Performance Tests
AWG2021 Service Manual
4-27
Step Display
Figure 4-11: MODE Menu Autostep Setting
H
Check that the DMM reading is in the range from +2.375 to +2.625 V
(100% modulation).
H
Set the function generator offset value to 0 V. Check that the DMM
reading is in the range from +1.125 to +1.375 V (50% modulation).
H
Set the function generator offset value to –1 V. Check that the DMM
voltage reading is in the range from –0.125 to +0.125 V
(0% modulation).
6. End procedure: Keep the test connections and instrument settings for the
next check.
Summary of Contents for Tektronix AWG2021
Page 4: ......
Page 16: ...Service Safety Summary xii AWG2021 Service Manual ...
Page 21: ...Specifications ...
Page 22: ......
Page 39: ...Operating Information ...
Page 40: ......
Page 60: ...Operating Instructions 2 20 AWG2021 Service Manual ...
Page 61: ...Theory of Operation ...
Page 62: ......
Page 66: ...Theory of Operation 3 4 AWG2021 Service Manual ...
Page 67: ...Performance Verification ...
Page 68: ......
Page 123: ...Adjustment Procedures ...
Page 124: ......
Page 135: ...Maintenance ...
Page 136: ......
Page 144: ...Maintenance 6 8 AWG2021 Service Manual ...
Page 188: ...Removal and Installation Procedures 6 52 AWG2021 Service Manual ...
Page 196: ...Troubleshooting 6 60 AWG2021 Service Manual TP100 5V TP101 GND Figure 6 34 Key Board ...
Page 201: ...Options ...
Page 202: ......
Page 222: ...Options and Accessories 7 20 AWG2021 Service Manual ...
Page 223: ...Replaceable Electrical Parts ...
Page 224: ......
Page 226: ...Replaceable Electrical Parts 8 2 AWG2021 Service Manual ...
Page 227: ...Diagrams ...
Page 228: ......
Page 230: ...Diagrams 9 2 AWG2021 Service Manual ...
Page 236: ...9 8 AWG2021 Service Manual ...
Page 237: ...Replaceable Mechanical Parts ...
Page 238: ......
Page 255: ...Replaceable Parts AWG2021 Service Manual 10 17 2 5 1 3 4 A9 2 Figure 10 7 Option 03 ...
Page 257: ...Replaceable Parts AWG2021 Service Manual 10 19 3 1 2 A25 4 Figure 10 8 Option 04 ...
Page 261: ......
Page 262: ......