Operating Instructions
2-10
AWG2021 Service Manual
The AWG2021 performs internal calibration routines using the Diag/Cal item in
the UTILITY menu, which check internal circuit function and adjust calibration
constants. Run these calibration routines whenever the AWG2021 undergoes a
temperature change. For instance, run the calibration routines after the
AWG2021 warms up at power-on. This warmup period and subsequent
calibration assures AWG2021 operation at optimum performance levels.
User Interface
The AWG2021 uses a combination of front-panel buttons, keys, a knob, and
on-screen menus to control generator functions. Some front-panel controls select
menus and manipulate menu items. Others enter values and units, allow manual
triggering, and turn on/off AWG2021 output. On-screen graphics show various
aspects of the current AWG2021 configuration.
On-screen menus set all AWG2021 functions except manual triggering and
output control. Main menus provide access to lower-level nested submenus.
Buttons in the center of the front panel select the main menus.
When you select a menu, the display shows the items controlled by that menu
and numeric values currently in effect. Buttons around the display select
lower-level menus, change menu selections, modify numeric values and units,
and execute functions.
Display
Figure 2–2 contains two examples of AWG2021 displays. To see the first display
on the AWG2021, press Edit in the MENU column; then press New Waveform
in the side menu. The second display shows an example of the message area.
Text after the illustration describes each display feature.
The status of the AWG2021 always appears in the status line. Status information
includes the interface status (refer to programmer manual), operating mode
status, trigger status (refer to user manual), and busy icon (a clock) which
appears while loading or saving waveform or sequence files.
This area displays the date and time set in the UTILITY menu. The date and time
display can be set to on or off.
If an error occurs during operation, this area displays an error message.
This area displays a label that matches the bottom menu that was selected.
Status Area
Date & Time Display
Error Display Area
Side MenuLabel
Summary of Contents for Tektronix AWG2021
Page 4: ......
Page 16: ...Service Safety Summary xii AWG2021 Service Manual ...
Page 21: ...Specifications ...
Page 22: ......
Page 39: ...Operating Information ...
Page 40: ......
Page 60: ...Operating Instructions 2 20 AWG2021 Service Manual ...
Page 61: ...Theory of Operation ...
Page 62: ......
Page 66: ...Theory of Operation 3 4 AWG2021 Service Manual ...
Page 67: ...Performance Verification ...
Page 68: ......
Page 123: ...Adjustment Procedures ...
Page 124: ......
Page 135: ...Maintenance ...
Page 136: ......
Page 144: ...Maintenance 6 8 AWG2021 Service Manual ...
Page 188: ...Removal and Installation Procedures 6 52 AWG2021 Service Manual ...
Page 196: ...Troubleshooting 6 60 AWG2021 Service Manual TP100 5V TP101 GND Figure 6 34 Key Board ...
Page 201: ...Options ...
Page 202: ......
Page 222: ...Options and Accessories 7 20 AWG2021 Service Manual ...
Page 223: ...Replaceable Electrical Parts ...
Page 224: ......
Page 226: ...Replaceable Electrical Parts 8 2 AWG2021 Service Manual ...
Page 227: ...Diagrams ...
Page 228: ......
Page 230: ...Diagrams 9 2 AWG2021 Service Manual ...
Page 236: ...9 8 AWG2021 Service Manual ...
Page 237: ...Replaceable Mechanical Parts ...
Page 238: ......
Page 255: ...Replaceable Parts AWG2021 Service Manual 10 17 2 5 1 3 4 A9 2 Figure 10 7 Option 03 ...
Page 257: ...Replaceable Parts AWG2021 Service Manual 10 19 3 1 2 A25 4 Figure 10 8 Option 04 ...
Page 261: ......
Page 262: ......