Preparation for Use
2-6
AWG2021 Service Manual
Applying and Interrupting Power
Consider the following information when you power on or power off the
AWG2021, or when external power loss occurs.
At power-on, the start-up diagnostics check the AWG2021 operation. The
start-up diagnostics take about 30 seconds. If all diagnostic items complete
without error, the AWG2021 displays PASS on the screen and then changes to the
SETUP menu.
If the diagnostics detect an error, the AWG2021 displays FAIL and the error
code. To exit the diagnostics menu, press any key; then the system displays the
SETUP menu. See section 6, Maintenance, for information on diagnostics and
fault isolation.
NOTE
. If the ambient temperature goes outside the specified operating tempera-
ture range, an error occurs during the diagnostics at power-on. If this happens,
power off the AWG2021 and wait until the chassis temperature is appropriate;
then switch the power on again.
CAUTION.
To prevent loss of internally stored adjustment constants, DO NOT
power off the AWG2021 when doing the internal calibration described in
section 5, Adjustment Procedures.
Wait for the AWG2021 to finish the operation when doing internal calibration or
saving waveform or sequence files. Improper power-off or unexpected loss of
power to the AWG2021 can result in the corruption of data stored in nonvolatile
memory.
PowerĆon Cycle
PowerĆoff Cycle
Summary of Contents for Tektronix AWG2021
Page 4: ......
Page 16: ...Service Safety Summary xii AWG2021 Service Manual ...
Page 21: ...Specifications ...
Page 22: ......
Page 39: ...Operating Information ...
Page 40: ......
Page 60: ...Operating Instructions 2 20 AWG2021 Service Manual ...
Page 61: ...Theory of Operation ...
Page 62: ......
Page 66: ...Theory of Operation 3 4 AWG2021 Service Manual ...
Page 67: ...Performance Verification ...
Page 68: ......
Page 123: ...Adjustment Procedures ...
Page 124: ......
Page 135: ...Maintenance ...
Page 136: ......
Page 144: ...Maintenance 6 8 AWG2021 Service Manual ...
Page 188: ...Removal and Installation Procedures 6 52 AWG2021 Service Manual ...
Page 196: ...Troubleshooting 6 60 AWG2021 Service Manual TP100 5V TP101 GND Figure 6 34 Key Board ...
Page 201: ...Options ...
Page 202: ......
Page 222: ...Options and Accessories 7 20 AWG2021 Service Manual ...
Page 223: ...Replaceable Electrical Parts ...
Page 224: ......
Page 226: ...Replaceable Electrical Parts 8 2 AWG2021 Service Manual ...
Page 227: ...Diagrams ...
Page 228: ......
Page 230: ...Diagrams 9 2 AWG2021 Service Manual ...
Page 236: ...9 8 AWG2021 Service Manual ...
Page 237: ...Replaceable Mechanical Parts ...
Page 238: ......
Page 255: ...Replaceable Parts AWG2021 Service Manual 10 17 2 5 1 3 4 A9 2 Figure 10 7 Option 03 ...
Page 257: ...Replaceable Parts AWG2021 Service Manual 10 19 3 1 2 A25 4 Figure 10 8 Option 04 ...
Page 261: ......
Page 262: ......