Guidance and manufacturer’s declaration – electromagnetic immunity
The MCC-1000MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the MCC-1000MD should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the MCC-1000MD,
including cables, than the
recommended separation
distance calculated from the
equation appliance to the
frequency of the transmitter.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bands
c
3 Vrms
d
= 1.2
√
P
6 Vrms
150 kHz to 80 MHz
in ISM bands
c
6 Vrms