Operation
There is a choice of three menu items in the main menu: the data logger
for storing the test values, the test parameters for changing all
parameters relevant to the test and the setup for changing the device’s
parameters. The selection is carried out by turning the rotary knob.
Press it to open the respective menu. Press the F3 function key to
return the unit to the check mode.
The selected menu item is indicated by a dot in the selection circle.
Menu
ESC
Data Logger
Test Parameter
Setup
4.6
Data logger
General information
Ultrasound levels and temperature test values can be stored in the data
logger. The data record includes the date, the tester (maximum eight
characters), the test site (maximum 16 characters), the type of probe
used, the ultrasound level and the temperature. There are two kinds of
tests:
1. Long-time test:
Storing ultrasonic and temperature test values
over a longer period at constant time intervals.
A maximum of 250 Long-time tests with
73 datasets each, or 1 Long-time test with
20,000 datasets can be recorded. 1 record
consists of the ultrasonic test value and the
temperature test value.
Selection