EN
| 45
9. Troubleshooting
Finding
Machine produces cooking
noises during heating
Specified temperature
is exceeded
No call-up
of the program possible
PRE
SET
LED indicator off
Possible cause
• No circulation of the cleaning liquid
• Temperature sensor does not measure the
average temperature (no circulation)
• Ultrasound energy continues to heat the liquid
(physical process)
• Execution of a manually set program
• Machine is switched off
Resolution
• e.g. also switch on ultrasound
• Circulate liquid manually or with ultrasound
• Only switch on ultrasound for a short time
• Press Start / Stop button briefly; then
it is possible to call up a program
• Switch on machine with on/off button