Appendix A: Schematic and Parts List
14
SiteStand Service Manual
A.4
SiteStand Wiring Diagram
IEC Input AC Receptacle (male)
Input Fuses:
Hot & Neutral
6.3A
B/W Printer Power
Output - IEC
Receptacle (female)
Auxiliary Power
Output - IEC
Receptacle (female)
100-240VAC
INPUT
Universal AC Input (100-240V 50-60Hz) Power
Adapter.
Output = 16VDC
CSA/UL approved to IEC601 Based Standards
Docking Stand PCB
16VDC
16VDC RTN
1: V+
2: V-
3: Video
4: GND
5: Print Control
6: Print Status
7: GND
8: Rx
9: Tx
10: GND
11: Rts
12: Cts
13: Sense
14: Spare
NOTE:
Docking Stand is disabled with
no voltages at 14 output
contacts until SonoSite 180 is
placed in Stand connecting V- to
Ground.
SonoSite 180
Printer Control / Status
Video Out to B/W Printer
Auxiliary Video Output
RS-232 Communications Port
Summary of Contents for SiteStand
Page 1: ...SonoSite SiteStand Mobile Docking Station Service Manual ...
Page 4: ... iv ...
Page 10: ...Chapter 1 Introduction 4 SiteStand Service Manual ...
Page 22: ...Appendix A Schematic and Parts List 16 SiteStand Service Manual A 6 SiteStand Assembly ...
Page 23: ...SiteStand Sleeve Assembly SiteStand Service Manual 17 A 7 SiteStand Sleeve Assembly ...