SiteStand Performance Tests
SiteStand Service Manual
11
To replace the electrical fuse:
1.
Disconnect the AC line cord from the hospital-grade electrical outlet.
2.
Disconnect the AC line cord from the bottom of the SiteStand mobile docking
station.
3.
Use a small, slotted screwdriver to remove the fuse drawer (located directly
below the input AC receptacle on the SiteStand mobile docking station).
CAUTION: You should identify the cause of the blown fuse prior to plugging
the docking station back into AC power.
4.
Carefully replace the blown fuse with a 250 volt, 6.3 amp, 5.0 x 20 mm Slo-
Blo®-type fuse.
5.
Install the fuse drawer.
6.
Connect the AC power to the SiteStand mobile docking station.
3.4
SiteStand Performance Tests
3.4.1
Setup
•
Install a SonoSite 180 or SonoHeart, without a battery into the docking station
•
Apply AC power, 120V 60Hz to SiteStand “AC in” receptacle
3.4.2
AC Power Verification
Use an AC Tester to verify both IEC output receptacles are wired correctly.
3.4.3
Docking Video
•
Turn on the system.
•
Connect a Test Monitor to each of the video outputs on the docking stand.
•
Verify the same video pattern exists on the Test Monitor as is present on the
System LCD Display.
3.4.4
RS-232
The RS-232 interface can only be field tested using the SiteLink Image Management
software. If that software is installed, verify the information on the PC according to
the SiteLink Image Management User Guide on the SiteLink CD-ROM.
3.4.5
Printer Interface
1.
Connect the black and white video printer to the SiteStand. Ensure that the
control and AC power cables are attached, but not the video cable.
2.
Verify that the printer produces an audible beep and that the user interface
replies with its own tone.
Summary of Contents for SiteStand
Page 1: ...SonoSite SiteStand Mobile Docking Station Service Manual ...
Page 4: ... iv ...
Page 10: ...Chapter 1 Introduction 4 SiteStand Service Manual ...
Page 22: ...Appendix A Schematic and Parts List 16 SiteStand Service Manual A 6 SiteStand Assembly ...
Page 23: ...SiteStand Sleeve Assembly SiteStand Service Manual 17 A 7 SiteStand Sleeve Assembly ...