Initial Issue
Issue Date 25/06/2010
User Manual: 8142/8182 Transceivers
08 Operational Test
Page
4of 6
15.
Un-tick the box marked “Update MPDB”
16.
Click on the ‘’Get Serial Number From HPT’’ this will display the serial number in the Serial Number
window.
17.
Click on Perform Test.
18.
An admittance test is performed on the unit as part of the tests. Admittance test windows will appear
on the screen during the test. The results will be flagged red, orange or green. See below
HPT Admittance Test Window
Screen Shot 4
19.
If any of the flags appear red, when prompted click NO for a successful test. You should now have the
screen shown in screen shot 5, note this screen shot indicates a system without a functioning Ethernet
connection.
Summary of Contents for 8142-000-01
Page 100: ......