![SOMNOmedics SOMNOscreen plus Instruction Manual Download Page 36](http://html1.mh-extra.com/html/somnomedics/somnoscreen-plus/somnoscreen-plus_instruction-manual_1318487036.webp)
- 36 -
5.10 ECG
a)
SOMNOscreen™ plus PSG+
Connect the two ECG cables to the plugs labelled
ECG
on the headbox.
Attach the yellow coloured electrode to a disposable electrode and secure it to ICR 4 (Fig. 5-26). The
green coloured electrode has to be fitted to ICR 2 parasternal left with a disposable electrode too.
Fig. 5-26: Attaching the ECG electrodes -
SOMNOscreen™ plus PSG+
Note:
Do not attach the
SOMNOscreen™ plus directly on the patient’s skin. Attach the SOMNOscreen™
plus over a pyjama or t-shirt.
Fig. 5-27: Connecting the EXG adaptor to the socket
Fig. 5-27 shows how to connect the black-grey coloured plug to the
SOMNOscreen™ plus.