41
Note:
This SOMNOmedics device complies with protection class IP 22 regarding ingression
of humidity and water. Cleaning should be performed with a lint-free and damp cloth.
Note:
Ensure that no liquids seep into the device during cleaning!
Note:
Other surface disinfectants listed by the DGHM are permitted as long as the surface of
the device or sensor will not be damaged.
Note:
The device and the sensors cannot be sterilized or autoclaved.
Disinfection of the sensors, electrodes, effort belts:
Item
Disinfectant
(Trade name)
Concentration Duration Frequency Other
Headbox
Wipe disinfection:
e. g.
Terralin Liquid
Mikrozid AF cloths
Ready to use
Ready to use
5 min
5 min
After use
Remove adhesive
residues if necessary.
Follow manufacturer’s
instructions!
Thermistor
Wipe disinfection:
e. g.
Terralin Liquid
Mikrozid AF cloths
Ready to use
Ready to use
5 min
5 min
After use
Remove adhesive
residues if necessary.
Follow manufacturer’s
instructions!