76
J-C-D
SommerXF starts with BOM
The BOM (Byte Order Mark) labels the downloaded data file to indicate that special characters within
the file are coded.
Setting
Description
on (default)
BOM is returned.
off
BOM is not returned.
J-C-E
Exposure lock
Sets data visibility on the MRL-7 display.
Setting
Description
off (default)
Measurement data are displayed.
display
Measurement data are displayed only after access code has been entered
(see Section
).
d data
Measurement data are displayed only after access code has been entered.
Additionally, downloading data to a connected flash drive requires the ac-
cess code (see Section
!
Attention
As long as the display is active, the entered access code keeps all system set-
tings of the logger unlocked!
J-C-F
LCD, Contrast
The contrast of the LCD-display.
Setting
Description
weak
low contrast
base (default)
base contrast
medium
medium contrast
high
high contrast
J-C-G
LCD, Brightness
The MRL-7 is equipped with an integrated ambient light sensor that is used to automatically adjust the
display brightness.
If the light intensity is below the specified value, the backlight brightness is automatically adjusted to
a suitable level. If the value is exceeded, the display backlight is switched off.
Value range
Default
Unit
0…3000
40
lm
J-C-H
Wind speed measurement duration
The time for measuring the wind speed with a connected anemometer.
Value range
Default
Unit
500…2000
1000
ms