_____________________________________________________
The TCS-110 is operated via TCPIP Ethernet; its default settings are
as follows:
IP Address:
192.168.1.88
Subnet Mask
255.255.255.0
Gateway:
192.168.1.1
IP Configuration
Static
Password
admin
Port Number
10001
4.1
S
TART
-
UP WHEN CONNECTED DIRECTLY TO A
PC:
1)
Plug the 24VDC power supply into your AC wall outlet or connect
24VDC to the TCS-110 as per section 3.2
2)
Connect your CAT-5 Ethernet Cross-Over Cable to the TCS-110 RJ-
45 ethernet jack. Set your computer to a static IP address in this
range: 192.168.1.2-254 excluding 88.
3)
Set your Subnet Mask to 255.255.255.0 and Gateway to 192.168.1.1.
4)
The TCS-110 is now ready for operation. See section 4.3.
4.2
S
TART
-UP
WHEN CONNECTED TO A
PC
THROUGH A
R
OUTER OR A
S
ERVER
1.
Plug the 24VDC power brick into your AC wall outlet or connect 24VDC to the TCS-110 as per
section 3.2.
2.
Connect a standard CAT-5 Ethernet cable from the computer to the router, and another standard
CAT-5 Ethernet cable from the router to the TCS-110 RJ-45 Ethernet jack.
3.
On the final test sheet provided upon shipment of the TCS-110 unit, find the IP address that has
been assigned to the unit. This address can be used in Section 4.3 to connect with the graphical
user interface.
4.3
CONNECTING
WITH
SSCS
GRAPHICAL
USER
INTERFACE
Solid State Cooling has developed a graphical user interface, or GUI, to be used to communicate with
the TCS-110 easily. For more information on the GUI and how to acquire it, contact the Sales
department at Solid State Cooling. If you do not wish to use this GUI, proceed to Section 5. (Note:
Establishing an Ethernet connection requires the use of the GUI.)
The following is an explanation of how to use the GUI to connect to the TCS-110 only after executing
either Section 4.1 or Section 4.2.
CAUTION
Do Not Change the
UART Settings. The
TCS-110 will not