Chapter 12 - Troubleshooting
PROFIBUS Tester BC-700-PB - User Manual
45
12
Troubleshooting
Problem
Causes and remedies
The display of the
PROFIBUS Tester 5
remains blank.
Possible cause:
The PROFIBUS Tester 5 either needs an external power
supply or you have to install the built-in battery.
Remedy:
Use the supplied external AC adapter to connect the tool to
the mains power supply.
Charge the built-in battery
Alternative cause:
The tool is defective.
The rechargeable battery is defective
Remedy:
Return the tool to the manufacturer.
Replace the rechargeable battery
The display shows the
following error
message: "USB
ERROR - Refer to
manual or disconn. for
stand-alone mode".
Possible cause:
This error message can be caused by poor physical
connection via USB ("loose" contact).
Remedy:
Check the USB cable and connector.
Alternative cause:
The device may be connected via external USB hub,
notebook docking stations or USB 3.0 port which could
evoke problems during driver installation
Remedy:
It is recommended to connect the unit directly to a USB 2.0
port
The baud rate is not
detected automatically
on a live PROFIBUS.
Possible cause:
Massive disturbances in the bus physics.
Remedy:
Manually set the baud rate and repeat the test.
A main menu function
is grayed out or is not
visible respectively
Possible cause:
Missing license
Out-dated firmware version is in use
Remedy:
Please check your bill of delivery if the license is included in
your order.
Please download the latest firmware version