Chapter 3: Controls and Features
Veterinary 3 Parameter Advisor
®
Operation Manual
3-5
Side Panel
The left side panel of your monitor contains all of the patient connector receptacles and the AC power
receptacle.
CONNECTOR
DESCRIPTION
INSTRUCTION
u
Oximetery
Connector (SpO
2
)
The SPO
2
parameter box will appear on the display when the patient
connector is attached to the monitor. Measured values for oxygen
saturation (%SpO
2
) in the blood and peripheral pulse rate (PPR) will be
displayed when the sensor is attached to the patient.
v
ECG Connector
The ECG parameter box will always appear on the display. A measured
value for the ECG heart rate (HR) will be displayed when the ECG leads
are attached to the patient and the ECG MONITOR menu option is
turned on. If five-lead processing is installed on your monitor, you
must also set the ECG LEADS PROCESSING menu option to five-lead
processing. See
Parameter Options
in
Chapter 4: Setting Up the Monitor
for instructions for turning on ECG MONITOR and setting ECG LEADS
PROCESSING.
CAUTION! A three-lead ECG cable must be used if 3-Lead
processing is selected. A five-lead ECG cable must
be used if 5-Lead processing is selected. Using the
incorrect cable for the selected mode might lead to
a floating reference or additional noise on the ECG
signal.
w
Non-Invasive
Blood Pressure
Connector (NIBP)
The NIBP parameter box will always appear on the display. Measured
values for non-invasive blood pressure (systolic, diastolic, and
mean) will be displayed when the most recent NIBP measurement is
complete.
ECG
NIBP
BP2
BP1
TEMP
SPO2
w
2
1
2
100-240AC
0.8A 50-60 Hz
F 1.6A/250V
REPLACE FUSE AS MARKED
RISK OF FIRE
WARNING:
CAUTION
u
y
w
v
Figure 3.4: Side Panel
x