Smiths Medical CADD-legacy Plus 6500 Quick Reference Card Download Page 6

Fluid is not flowing from the reservoir to the 

pump. Check for a kink or air bubble in the 

tubing between the reservoir and pump. Remove 

the occlusion to resume operation or press 

STOP/START or NEXT to stop the pump and 

silence alarm for two minutes, then remove the 

occlusion and restart the pump.
The cassette is not aligned with the pump, or is 

damaged, or a malfunction of the pump sensor(s) 

is occuring. Reposition the pump to silence the 

alarm. If repositioning the pump does not silence 

the alarm within two minutes, the pump will 

display, “No Disposable, Clamp Tubing.”
Air in fluid path. Press NEXT or STOP/START to 

silence alarm. Make sure tubing is threaded in 

air detector properly. If the fluid path contains 

air, close clamps, disconnect from patient and 

prime fluid path.
Pump is stopped. Start pump, if appropriate.

Verify all programming screens by pressing  

ENTER/CLEAR before moving to next screen or 

starting pump.
A value was not saved by pressing ENTER/

CLEAR. Press NEXT to resume programming.
An error has occurred. Remove pump from 

service.
The pump was running when batteries were 

removed or, the batteries were removed within  

15 seconds after stopping the pump. Batteries 

must be reinstalled.

imPorTAnT: Always stop pump before 

removing batteries.
The disposable was removed, or the cassette is 

not aligned with the pump or is damaged, or a  

malfunction of the pump sensor(s) occurring.  

Clamp the tubing immediately. Press STOP/

START or NEXT to silence alarm.
The pump is due for service.

AlArms & TroubleshooTing 

Continued

The CADD-legacy

®

 Plus pump signals with audible alarms and messages 

that appear on the display when a condition requires your attention.

upstream  

occlusion
   Two-tone alarm

Air in line  

Detected
   Two-tone alarm

stopped

   Three beeps 

every five minutes
Programming 

incomplete

Value not saved

error
   Two-tone alarm

no message  

Displayed
   Two-tone alarm

no Disposable, 

Clamp Tubing
   Two-beep  
(long-short) alarm
service Due
   Two-tone alarm

screen Displays  

Current Pump 

status
   Two-beep  
(long-short) alarm

Warning: This Quick reference Card should be used by clinicians 

only. Do not permit patients to have access to this card, as the 

information would allow access to all programming and operating 

functions. Assistance with the CADD-legacy

®

 Plus pump is 

available to clinicians 24-hours-a-day by calling 1-800-258-5361.

CADD, CADD-Legacy, the Medication Cassette Reservoir design and the Smiths Medical 

and CADD design marks are trademarks of Smiths Medical. The symbol ® indicates 

the trademark is registered in the U.S. Patent and Trademark Office and certain other 

countries. ©2010 Smiths Medical. All rights reserved.    11/10    IN19892

smiths medical AsD, inc.

St. Paul, MN 55112, USA
Phone: 1-214-618-0218
Toll-Free USA: 1-800-258-5361

www.smiths-medical.com

smiths medical is part of the global technology business smiths group plc.

eC Authorized representative

Smiths Medical international Ltd.

TN25 4BF, UK
Phone: +44 (0) 1233 722100

Product(s) described may not be licensed or available for sale in Canada or other  

countries outside of the United States.

smiths medical Canada ltd.

Markham, Ontario, Canada, L3R 4Y8
Phone: 905-477-2000
Toll-Free: 1-800-387-4346

Summary of Contents for CADD-legacy Plus 6500

Page 1: ...Adapter back of pump h Battery Compartment back of pump i Cassette Latch attaches cassette to pump side of pump KeYPAD STOP START Stops and starts the infusion ENTER CLEAR Enters or clears displayed value PRIME Fills tubing with fluid LOCK Displays or changes Lock Level security level NEXT Advances to next programming screen Increases or decreases displayed values or scrolls through menu items ON ...

Page 2: ...ibed will cause over or under delivery of medication 7 Press ENTER CLEAR to clear the given value The display will show 0 00 8 Press NEXT 9 Verify the setting is correct To change setting see Biomed Functions section 10 Press NEXT 11 Verify the setting is correct To change setting see Biomed Functions section 12 Press NEXT 13 Press NEXT repeatedly to review program inTermiTTenT DeliVerY batteries ...

Page 3: ...ntil pump beeps and powers up 1 Press and hold ON OFF until appears on the display 2 Release ON OFF key 3 The screen will go blank as the pump goes into a lower power state 1 Stop the pump 2 Push down and hold the arrow button on the battery door while sliding the door off Remove and discard old batteries imPorTAnT Alwaysstoppumpbeforeremovingbatteries 3 Install new batteries matching polarities s...

Page 4: ...ctions To Access To Turn up stream sensor on oFF To Change Delivery mode To exit biomed Functions Pump must be stopped and in ll0 1 Press LOCK 2 Press LOCK or ENTER CLEAR 3 Press orÎ until the biomed code appears 4 Press LOCK or ENTER CLEAR 5 Press NEXT to go through settings 6 Press NEXT until AIR DETECTOR appears 7 Press orÎ until desired setting appears 8 Press ENTER CLEAR 9 Press NEXT WArning ...

Page 5: ...e and discard used cassette 1 Close all tubing clamps 2 Insertnewcassettehooksintopump shingepins 3 Place pump with cassette on firm surface and push down on top of pump so cassette fits tightly against the pump 4 Insert a coin into the lock push in and turn until the line on the lock lines up with the arrow on the side of the pump 5 Gently twist and pull on cassette to make sure it is attached pr...

Page 6: ...e pump is due for service AlArms TroubleshooTing Continued TheCADD legacy Pluspumpsignalswithaudiblealarmsandmessages thatappearonthedisplaywhenaconditionrequiresyourattention upstream occlusion Two tone alarm Air in line Detected Two tone alarm stopped Three beeps every five minutes Programming incomplete Value not saved error Two tone alarm no message Displayed Two tone alarm no Disposable Clamp...

Reviews: