Fluid is not flowing from the reservoir to the
pump. Check for a kink or air bubble in the
tubing between the reservoir and pump. Remove
the occlusion to resume operation or press
STOP/START or NEXT to stop the pump and
silence alarm for two minutes, then remove the
occlusion and restart the pump.
The cassette is not aligned with the pump, or is
damaged, or a malfunction of the pump sensor(s)
is occuring. Reposition the pump to silence the
alarm. If repositioning the pump does not silence
the alarm within two minutes, the pump will
display, “No Disposable, Clamp Tubing.”
Air in fluid path. Press NEXT or STOP/START to
silence alarm. Make sure tubing is threaded in
air detector properly. If the fluid path contains
air, close clamps, disconnect from patient and
prime fluid path.
Pump is stopped. Start pump, if appropriate.
Verify all programming screens by pressing
ENTER/CLEAR before moving to next screen or
starting pump.
A value was not saved by pressing ENTER/
CLEAR. Press NEXT to resume programming.
An error has occurred. Remove pump from
service.
The pump was running when batteries were
removed or, the batteries were removed within
15 seconds after stopping the pump. Batteries
must be reinstalled.
imPorTAnT: Always stop pump before
removing batteries.
The disposable was removed, or the cassette is
not aligned with the pump or is damaged, or a
malfunction of the pump sensor(s) occurring.
Clamp the tubing immediately. Press STOP/
START or NEXT to silence alarm.
The pump is due for service.
AlArms & TroubleshooTing
Continued
The CADD-legacy
®
Plus pump signals with audible alarms and messages
that appear on the display when a condition requires your attention.
upstream
occlusion
Two-tone alarm
Air in line
Detected
Two-tone alarm
stopped
Three beeps
every five minutes
Programming
incomplete
Value not saved
error
Two-tone alarm
no message
Displayed
Two-tone alarm
no Disposable,
Clamp Tubing
Two-beep
(long-short) alarm
service Due
Two-tone alarm
screen Displays
Current Pump
status
Two-beep
(long-short) alarm
Warning: This Quick reference Card should be used by clinicians
only. Do not permit patients to have access to this card, as the
information would allow access to all programming and operating
functions. Assistance with the CADD-legacy
®
Plus pump is
available to clinicians 24-hours-a-day by calling 1-800-258-5361.
CADD, CADD-Legacy, the Medication Cassette Reservoir design and the Smiths Medical
and CADD design marks are trademarks of Smiths Medical. The symbol ® indicates
the trademark is registered in the U.S. Patent and Trademark Office and certain other
countries. ©2010 Smiths Medical. All rights reserved. 11/10 IN19892
smiths medical AsD, inc.
St. Paul, MN 55112, USA
Phone: 1-214-618-0218
Toll-Free USA: 1-800-258-5361
www.smiths-medical.com
smiths medical is part of the global technology business smiths group plc.
eC Authorized representative
Smiths Medical international Ltd.
TN25 4BF, UK
Phone: +44 (0) 1233 722100
Product(s) described may not be licensed or available for sale in Canada or other
countries outside of the United States.
smiths medical Canada ltd.
Markham, Ontario, Canada, L3R 4Y8
Phone: 905-477-2000
Toll-Free: 1-800-387-4346