Smiths Medical CADD-legacy Plus 6500 Quick Reference Card Download Page 5

AlArms & TroubleshooTing

ChAnging The CAsseTTe

To remove used

Cassette

To Attach new

Cassette

The CADD-legacy

®

 Plus pump signals with audible alarms and messages 

that appear on the display when a condition requires your attention.

Level of fluid in the reservoir is low. Prepare to 

install new reservoir.
Reservoir volume has reached 0.0 ml. Press 

NEXT or STOP/START to silence alarm. Install a 
new reservoir, if appropriate.
Batteries are low, but pump is operable. 
Change batteries soon.

Batteries are depleted and cannot operate 
pump; install new batteries immediately. 
With AC adapter attached, batteries have been 
removed while pump was running, or you have 
tried to start pump with depleted batteries. 
Install new batteries.
High pressure is being caused by an occlusion in 
the fluid path between the pump and the patient. 
Remove the occlusion to resume operation.  
Or press STOP/START or NEXT to stop the  
pump and silence the alarm for two minutes, 
then remove occlusion and restart pump.

res Vol low
   Three beeps
reservoir  

Volume empty
   Two-tone alarm
low bat
   3 two-tone beeps  

every five minutes 

battery Depleted
   Two-tone alarm

battery removed 

Pump Won’t run
   Two-tone alarm

high Pressure
   Two-tone alarm

1.  Stop the pump.
2.  Close all tubing clamps.
3.  Disconnect tubing from patient.
4.  Use coin to unlatch used cassette.
5.  Remove and discard used cassette.
1.   Close all tubing clamps.
2.   Insert new cassette hooks into pump’s hinge pins.
3.   Place pump with cassette on firm surface 

and push down on top of pump so cassette 
fits tightly against the pump.

4.   Insert a coin into the lock, push in and turn 

until the line on the lock lines up with the 
arrow on the side of the pump.

5.  Gently twist and pull on cassette to make 

sure it is attached properly.

6.  Press ENTER/CLEAR to reset reservoir 

volume, if appropriate.

7.  If the Air Detector is in use, make a small 

loop of tubing underneath the air detector 
and hold it with your thumb. Place the 
tubing over the groove in the air detector 
and tuck it under the catch. Seat the tubing 
into the groove by gently pulling the tube 
upward until it is under the retention nubs 
and flat in the groove.

WArning: Attach the cassette (the part of the 

medication Cassette reservoir or CADD

®

 

administration set that attaches to the pump) 

properly. An improperly attached or detached 

cassette could result in unregulated gravity  

infusion of medication from the fluid container 

or a reflux of blood. if you are using a CADD

®

 

administration set or medication Cassette 

reservoir that does not have the flow stop 

feature (catalog number does not start with 

21-73xx): You must use a CADD

®

 extension set 

with an integral anti-siphon valve or a CADD

®

 

administration set with either and integral 

or add on anti-siphon valve to protect against 

unregulated gravity infusion that can result 

from an improperly attached cassette.

Summary of Contents for CADD-legacy Plus 6500

Page 1: ...Adapter back of pump h Battery Compartment back of pump i Cassette Latch attaches cassette to pump side of pump KeYPAD STOP START Stops and starts the infusion ENTER CLEAR Enters or clears displayed value PRIME Fills tubing with fluid LOCK Displays or changes Lock Level security level NEXT Advances to next programming screen Increases or decreases displayed values or scrolls through menu items ON ...

Page 2: ...ibed will cause over or under delivery of medication 7 Press ENTER CLEAR to clear the given value The display will show 0 00 8 Press NEXT 9 Verify the setting is correct To change setting see Biomed Functions section 10 Press NEXT 11 Verify the setting is correct To change setting see Biomed Functions section 12 Press NEXT 13 Press NEXT repeatedly to review program inTermiTTenT DeliVerY batteries ...

Page 3: ...ntil pump beeps and powers up 1 Press and hold ON OFF until appears on the display 2 Release ON OFF key 3 The screen will go blank as the pump goes into a lower power state 1 Stop the pump 2 Push down and hold the arrow button on the battery door while sliding the door off Remove and discard old batteries imPorTAnT Alwaysstoppumpbeforeremovingbatteries 3 Install new batteries matching polarities s...

Page 4: ...ctions To Access To Turn up stream sensor on oFF To Change Delivery mode To exit biomed Functions Pump must be stopped and in ll0 1 Press LOCK 2 Press LOCK or ENTER CLEAR 3 Press orÎ until the biomed code appears 4 Press LOCK or ENTER CLEAR 5 Press NEXT to go through settings 6 Press NEXT until AIR DETECTOR appears 7 Press orÎ until desired setting appears 8 Press ENTER CLEAR 9 Press NEXT WArning ...

Page 5: ...e and discard used cassette 1 Close all tubing clamps 2 Insertnewcassettehooksintopump shingepins 3 Place pump with cassette on firm surface and push down on top of pump so cassette fits tightly against the pump 4 Insert a coin into the lock push in and turn until the line on the lock lines up with the arrow on the side of the pump 5 Gently twist and pull on cassette to make sure it is attached pr...

Page 6: ...e pump is due for service AlArms TroubleshooTing Continued TheCADD legacy Pluspumpsignalswithaudiblealarmsandmessages thatappearonthedisplaywhenaconditionrequiresyourattention upstream occlusion Two tone alarm Air in line Detected Two tone alarm stopped Three beeps every five minutes Programming incomplete Value not saved error Two tone alarm no message Displayed Two tone alarm no Disposable Clamp...

Reviews: