ProgrAmming / seT-uP seQuenCe
CONTINUED
PumP oPerATions
1. Stop the pump.
2. Press LOCK.
3. Press
´
or
Î
until desired lock level appears.
4. Press LOCK or ENTER/CLEAR.
5. Press
´
or
Î
until the lock level code appears.
6. Press LOCK or ENTER/CLEAR.
1. Press and hold STOP/START until
(– – – – – – – – – – – –) appears on the display.
2. Release STOP/START key. STOPPED will appear on
the display when the pump is stopped.
1. Press and hold STOP/START until
(– – – – – – – – – – – –) disappears from the display.
2. Release STOP/START key. RUN will appear on
the display when the pump is running.
Pump must be stopped and in ll0 or ll1.
WArning: Do not prime the fluid path with the tubing
connected to a patient as this could result in over
delivery of medication or air embolism.
1. Press and hold PRIME until the word PRIME
appears on the display, along with (– – – – – – – – – – – –).
2. Release the PRIME key.
3. Press and hold PRIME. Priming appears on the display.
Continue priming until the fluid path is free of air.
4. Press NEXT to return to the main screen.
1. Stop the pump.
2. Press NEXT to display the reservoir volume screen.
3. Press ENTER/CLEAR to reset the value to
previously programmed amount.
1. Press and hold ON/OFF until pump beeps and
powers up.
1. Press and hold ON/OFF until (•••• •••• ••••)
appears on the display.
2. Release ON/OFF key.
3. The screen will go blank as the pump goes into a
lower power state.
1. Stop the pump.
2. Push down and hold the arrow button on the battery
door while sliding the door off. Remove and discard
old batteries.
imPorTAnT: Always stop pump before removing batteries.
3. Install new batteries, matching polarities shown on
pump. Replace battery door and close.
4. Start the pump.
WArning: if a gap is present anywhere between
the battery door and the pump housing, the door is
not properly latched. if the battery door becomes
detached or loose, the batteries will not be properly
secured which could result in loss of power or non-
delivery of medication.
WArning: Do not use rechargeable niCad or nickel
metal hydride (nimh) batteries. Do not use carbon
zinc (“heavy duty”) batteries.
Change
the lock
level
stop the
Pump
start the
Pump
22. Verify the setting is correct. (To change
setting, see Biomed Functions section.)
23. Press NEXT.
24. Verify the setting is correct. (To change
setting, see Biomed Functions section.)
25. Press NEXT.
26. Press NEXT repeatedly to
review program.
Verify Air Detector
status
Verify upstream
sensor status
Prime the
Fluid Path
reset the
reservoir
Volume
Turn the
Pump on
Turn the
Pump off
Change the
batteries
Verify
Programming