ProgrAmming / seT-uP seQuenCe 

CONTINUED

PumP oPerATions

1.  Stop the pump.
2.  Press LOCK.
3.  Press 

´

or

Î

 until desired lock level appears.

4.  Press LOCK or ENTER/CLEAR.
5.  Press 

´

or

Î

 until the lock level code appears.

6.  Press LOCK or ENTER/CLEAR.
1.  Press and hold STOP/START until  

(– – – –  – – – –  – – – –) appears on the display.

2.  Release STOP/START key. STOPPED will appear on 

the display when the pump is stopped.

1.  Press and hold STOP/START until  

(– – – –  – – – –  – – – –) disappears from the display. 

2.  Release STOP/START key. RUN will appear on  

the display when the pump is running.

Pump must be stopped and in ll0 or ll1.
WArning: Do not prime the fluid path with the tubing  

connected to a patient as this could result in over 

delivery of medication or air embolism.

1.  Press and hold PRIME until the word PRIME 

appears on the display, along with (– – – –  – – – –  – – – –).

2.  Release the PRIME key.
3.  Press and hold PRIME. Priming appears on the display. 

Continue priming until the fluid path is free of air.

4.  Press NEXT to return to the main screen.
1.  Stop the pump.
2.  Press NEXT to display the reservoir volume screen.
3.  Press ENTER/CLEAR to reset the value to 

previously programmed amount.

1.  Press and hold ON/OFF until pump beeps and 

powers up.

1.  Press and hold ON/OFF until (•••• •••• ••••) 

appears on the display.

2.  Release ON/OFF key.
3.  The screen will go blank as the pump goes into a 

lower power state.

1.  Stop the pump.
2.  Push down and hold the arrow button on the battery 

door while sliding the door off. Remove and discard 

old batteries.

imPorTAnT: Always stop pump before removing batteries.

3.  Install new batteries, matching polarities shown on 

pump. Replace battery door and close. 

4.  Start the pump.

WArning: if a gap is present anywhere between 

the battery door and the pump housing, the door is 

not properly latched. if the battery door becomes 

detached or loose, the batteries will not be properly 

secured which could result in loss of power or non-

delivery of medication.

WArning: Do not use rechargeable niCad or nickel 

metal hydride (nimh) batteries. Do not use carbon 

zinc (“heavy duty”) batteries. 

Change 

the lock 

level

stop the 

Pump

start the 

Pump

22.  Verify the setting is correct. (To change
 

setting, see Biomed Functions section.)

23.  Press NEXT.
24.  Verify the setting is correct. (To change 

setting, see Biomed Functions section.) 

25.  Press NEXT.
26.  Press NEXT repeatedly to  

review program.

Verify Air Detector 

status

Verify upstream  

sensor status

Prime the 

Fluid Path

reset the 

reservoir 

Volume

Turn the 

Pump on

Turn the 

Pump off

Change the 

batteries

Verify  

Programming

Summary of Contents for CADD-legacy Plus 6500

Page 1: ...Adapter back of pump h Battery Compartment back of pump i Cassette Latch attaches cassette to pump side of pump KeYPAD STOP START Stops and starts the infusion ENTER CLEAR Enters or clears displayed value PRIME Fills tubing with fluid LOCK Displays or changes Lock Level security level NEXT Advances to next programming screen Increases or decreases displayed values or scrolls through menu items ON ...

Page 2: ...ibed will cause over or under delivery of medication 7 Press ENTER CLEAR to clear the given value The display will show 0 00 8 Press NEXT 9 Verify the setting is correct To change setting see Biomed Functions section 10 Press NEXT 11 Verify the setting is correct To change setting see Biomed Functions section 12 Press NEXT 13 Press NEXT repeatedly to review program inTermiTTenT DeliVerY batteries ...

Page 3: ...ntil pump beeps and powers up 1 Press and hold ON OFF until appears on the display 2 Release ON OFF key 3 The screen will go blank as the pump goes into a lower power state 1 Stop the pump 2 Push down and hold the arrow button on the battery door while sliding the door off Remove and discard old batteries imPorTAnT Alwaysstoppumpbeforeremovingbatteries 3 Install new batteries matching polarities s...

Page 4: ...ctions To Access To Turn up stream sensor on oFF To Change Delivery mode To exit biomed Functions Pump must be stopped and in ll0 1 Press LOCK 2 Press LOCK or ENTER CLEAR 3 Press orÎ until the biomed code appears 4 Press LOCK or ENTER CLEAR 5 Press NEXT to go through settings 6 Press NEXT until AIR DETECTOR appears 7 Press orÎ until desired setting appears 8 Press ENTER CLEAR 9 Press NEXT WArning ...

Page 5: ...e and discard used cassette 1 Close all tubing clamps 2 Insertnewcassettehooksintopump shingepins 3 Place pump with cassette on firm surface and push down on top of pump so cassette fits tightly against the pump 4 Insert a coin into the lock push in and turn until the line on the lock lines up with the arrow on the side of the pump 5 Gently twist and pull on cassette to make sure it is attached pr...

Page 6: ...e pump is due for service AlArms TroubleshooTing Continued TheCADD legacy Pluspumpsignalswithaudiblealarmsandmessages thatappearonthedisplaywhenaconditionrequiresyourattention upstream occlusion Two tone alarm Air in line Detected Two tone alarm stopped Three beeps every five minutes Programming incomplete Value not saved error Two tone alarm no message Displayed Two tone alarm no Disposable Clamp...

Reviews: