PumP oPerATions
CONTINUED
main screen
Displays status of pump (STOPPED or RUN)
ConTinuous moDe
reservoir Volume
Volume of fluid in reservoir
Continuous rate
Continuous rate of infusion (ml/hr)
given
Total amount of medication delivered since
display was last cleared
Air Detector
Indicates if air detector is turned on or off
(status only)
and high or low sensitivity level
upstream sensor
Indicates if turned on or off
(status only)
inTermiTTenT moDe
reservoir Volume
Volume of fluid in reservoir
Dose Volume
Volume of an individual dose
Dose Duration
Amount of time over which an individual
dose is infused
Dose Cycle
Time from start of one dose to start of next
dose
KVo rate
Keep Vein Open rate in ml/hr delivered
in between doses
Dose remaining
Time remaining for delivery of current dose
(only appears during dosing or if dose is
interrupted)
Dose starts in
Amount of time before start of next dose
given
Total amount of medication delivered since
display was last cleared
Air Detector
Indicates if air detector is turned on or off
(status only)
and high or low sensitivity level
upstream sensor
Indicates if turned on or off
(status only)
ProgrAmming sCreens
biomed
Functions
To Access
To Turn
up-
stream
sensor
on/oFF
To Change
Delivery
mode
To exit
biomed
Functions
Pump must be stopped and in ll0.
1. Press LOCK.
2. Press LOCK or ENTER/CLEAR.
3. Press
´
or
Î
until the biomed code appears.
4. Press LOCK or ENTER/CLEAR.
5. Press NEXT to go through settings.
6. Press NEXT until AIR DETECTOR appears.
7. Press
´
or
Î
until desired setting appears.
8. Press ENTER/CLEAR.
9. Press NEXT.
WArning: When the air detector is turned off, the
pump will not detect air in the fluid path. Periodically
inspect the fluid path and remove any air to prevent
air embolism.
10.Press NEXT until UPSTREAM SENSOR appears.
11.Press
´
or
Î
until desired setting appears.
12.Press ENTER/CLEAR.
13.Press NEXT.
WArning: When the upstream occlusion sensor
is turned off, the pump will not detect occlusions
upstream (between pump and reservoir). Periodically
inspect the reservoir for decreasing reservoir volume
and inspect the fluid path for kinks, a closed clamp, or
other upstream occlusion. upstream occlusions could
result in under- or non-delivery of medications.
14.Press NEXT until DELIVERY MODE appears
(Intermittent or Continuous).
15.Press
´
or
Î
until desired setting appears.
16.Press ENTER/CLEAR.
17. Press NEXT until NEXT FOR BIOMED, ENTER FOR
MAIN appears.
18.Press ENTER/CLEAR to return to main screen. Or
press NEXT to continue in Biomed settings loop.
To Turn Air
Detector
on-high,
on-loW
or oFF