Smiths Medical CADD-legacy Plus 6500 Quick Reference Card Download Page 4

PumP oPerATions 

CONTINUED

main screen 

Displays status of pump (STOPPED or RUN)

ConTinuous moDe
reservoir Volume
 

Volume of fluid in reservoir

Continuous rate 

Continuous rate of infusion (ml/hr)

given 

Total amount of medication delivered since  
display was last cleared

Air Detector 

Indicates if air detector is turned on or off

(status only)  

and high or low sensitivity level

upstream sensor  

Indicates if turned on or off

(status only
inTermiTTenT moDe
reservoir Volume
 

Volume of fluid in reservoir

Dose Volume 

Volume of an individual dose

Dose Duration 

Amount of time over which an individual  
dose is infused

Dose Cycle 

Time from start of one dose to start of next 
dose

KVo rate 

Keep Vein Open rate in ml/hr delivered  
in between doses

Dose remaining 

Time remaining for delivery of current dose 
(only appears during dosing or if dose is 
interrupted)

Dose starts in 

Amount of time before start of next dose

given 

Total amount of medication delivered since  
display was last cleared

Air Detector  

Indicates if air detector is turned on or off

(status only)  

and high or low sensitivity level

upstream sensor  

Indicates if turned on or off

(status only)

ProgrAmming sCreens

biomed 

Functions

To Access

To Turn 

up-

stream

sensor 

on/oFF

To Change 

Delivery

mode
To exit 

biomed

Functions

Pump must be stopped and in ll0.

1.  Press LOCK.
2.  Press LOCK or ENTER/CLEAR.
3. Press 

´

or

Î

 until the biomed code appears.

4.  Press LOCK or ENTER/CLEAR.
5.  Press NEXT to go through settings.
6.  Press NEXT until AIR DETECTOR appears.
7. Press 

´

or

Î

 until desired setting appears.

8.  Press ENTER/CLEAR.
9.  Press NEXT.

WArning: When the air detector is turned off, the 

pump will not detect air in the fluid path. Periodically 

inspect the fluid path and remove any air to prevent 

air embolism.
10.Press NEXT until UPSTREAM SENSOR appears.
11.Press 

´

or

Î

 until desired setting appears.

12.Press ENTER/CLEAR.
13.Press NEXT.

WArning: When the upstream occlusion sensor 

is turned off, the pump will not detect occlusions 

upstream (between pump and reservoir). Periodically 

inspect the reservoir for decreasing reservoir volume 

and inspect the fluid path for kinks, a closed clamp, or 

other upstream occlusion. upstream occlusions could 

result in under- or non-delivery of medications.
14.Press NEXT until DELIVERY MODE appears 

(Intermittent or Continuous).

15.Press 

´

or

Î

 until desired setting appears.

16.Press ENTER/CLEAR.
17. Press NEXT until NEXT FOR BIOMED, ENTER FOR 

MAIN appears.

18.Press ENTER/CLEAR to return to main screen. Or 

press NEXT to continue in Biomed settings loop. 

To Turn Air  

Detector  

on-high,  

on-loW  

or oFF

Summary of Contents for CADD-legacy Plus 6500

Page 1: ...Adapter back of pump h Battery Compartment back of pump i Cassette Latch attaches cassette to pump side of pump KeYPAD STOP START Stops and starts the infusion ENTER CLEAR Enters or clears displayed value PRIME Fills tubing with fluid LOCK Displays or changes Lock Level security level NEXT Advances to next programming screen Increases or decreases displayed values or scrolls through menu items ON ...

Page 2: ...ibed will cause over or under delivery of medication 7 Press ENTER CLEAR to clear the given value The display will show 0 00 8 Press NEXT 9 Verify the setting is correct To change setting see Biomed Functions section 10 Press NEXT 11 Verify the setting is correct To change setting see Biomed Functions section 12 Press NEXT 13 Press NEXT repeatedly to review program inTermiTTenT DeliVerY batteries ...

Page 3: ...ntil pump beeps and powers up 1 Press and hold ON OFF until appears on the display 2 Release ON OFF key 3 The screen will go blank as the pump goes into a lower power state 1 Stop the pump 2 Push down and hold the arrow button on the battery door while sliding the door off Remove and discard old batteries imPorTAnT Alwaysstoppumpbeforeremovingbatteries 3 Install new batteries matching polarities s...

Page 4: ...ctions To Access To Turn up stream sensor on oFF To Change Delivery mode To exit biomed Functions Pump must be stopped and in ll0 1 Press LOCK 2 Press LOCK or ENTER CLEAR 3 Press orÎ until the biomed code appears 4 Press LOCK or ENTER CLEAR 5 Press NEXT to go through settings 6 Press NEXT until AIR DETECTOR appears 7 Press orÎ until desired setting appears 8 Press ENTER CLEAR 9 Press NEXT WArning ...

Page 5: ...e and discard used cassette 1 Close all tubing clamps 2 Insertnewcassettehooksintopump shingepins 3 Place pump with cassette on firm surface and push down on top of pump so cassette fits tightly against the pump 4 Insert a coin into the lock push in and turn until the line on the lock lines up with the arrow on the side of the pump 5 Gently twist and pull on cassette to make sure it is attached pr...

Page 6: ...e pump is due for service AlArms TroubleshooTing Continued TheCADD legacy Pluspumpsignalswithaudiblealarmsandmessages thatappearonthedisplaywhenaconditionrequiresyourattention upstream occlusion Two tone alarm Air in line Detected Two tone alarm stopped Three beeps every five minutes Programming incomplete Value not saved error Two tone alarm no message Displayed Two tone alarm no Disposable Clamp...