Smiths Medical CADD-legacy Plus 6500 Quick Reference Card Download Page 1

Rate is in

ml/hr

®

®

A

b

C

D

e

F

g

h

i

CADD-legacy

®

 Plus

infusion Pump 

Model 6500

Quick Reference Card for Clinicians

lock level 0 (ll0)

A

 

Display shows programmed values  

and messages

b

 

AC Adapter Jack used to plug in AC Adapter

C

 

Accessory Jack (not currently in use)

D

 

Indicator Light indicates AC power in use

e

 

Air Detector

F

 

Cassette (part of reservoir or administration set  

that attaches to pump)

g

 

Threaded Mounting Hole for use with Polemount  

Bracket Adapter (back of pump)

h

 

Battery Compartment (back of pump)

i

 

Cassette Latch attaches cassette to pump (side of pump)

KeYPAD
STOP/START 

Stops and starts the infusion

ENTER/CLEAR 

Enters or clears displayed value

PRIME 

Fills tubing with fluid

LOCK 

Displays or changes Lock Level (security level)

NEXT 

Advances to next programming screen

   

Increases or decreases displayed values  

 

or scrolls through menu items

ON/OFF 

Turns the pump on or off (low power)

Read the entire Operator’s Manual before operating the CADD-Legacy

®

 

ambulatory infusion pump. Failure to properly follow warnings, cautions  

and instructions could result in death or serious injury to the patient.

´
Î

Summary of Contents for CADD-legacy Plus 6500

Page 1: ...Adapter back of pump h Battery Compartment back of pump i Cassette Latch attaches cassette to pump side of pump KeYPAD STOP START Stops and starts the infusion ENTER CLEAR Enters or clears displayed value PRIME Fills tubing with fluid LOCK Displays or changes Lock Level security level NEXT Advances to next programming screen Increases or decreases displayed values or scrolls through menu items ON ...

Page 2: ...ibed will cause over or under delivery of medication 7 Press ENTER CLEAR to clear the given value The display will show 0 00 8 Press NEXT 9 Verify the setting is correct To change setting see Biomed Functions section 10 Press NEXT 11 Verify the setting is correct To change setting see Biomed Functions section 12 Press NEXT 13 Press NEXT repeatedly to review program inTermiTTenT DeliVerY batteries ...

Page 3: ...ntil pump beeps and powers up 1 Press and hold ON OFF until appears on the display 2 Release ON OFF key 3 The screen will go blank as the pump goes into a lower power state 1 Stop the pump 2 Push down and hold the arrow button on the battery door while sliding the door off Remove and discard old batteries imPorTAnT Alwaysstoppumpbeforeremovingbatteries 3 Install new batteries matching polarities s...

Page 4: ...ctions To Access To Turn up stream sensor on oFF To Change Delivery mode To exit biomed Functions Pump must be stopped and in ll0 1 Press LOCK 2 Press LOCK or ENTER CLEAR 3 Press orÎ until the biomed code appears 4 Press LOCK or ENTER CLEAR 5 Press NEXT to go through settings 6 Press NEXT until AIR DETECTOR appears 7 Press orÎ until desired setting appears 8 Press ENTER CLEAR 9 Press NEXT WArning ...

Page 5: ...e and discard used cassette 1 Close all tubing clamps 2 Insertnewcassettehooksintopump shingepins 3 Place pump with cassette on firm surface and push down on top of pump so cassette fits tightly against the pump 4 Insert a coin into the lock push in and turn until the line on the lock lines up with the arrow on the side of the pump 5 Gently twist and pull on cassette to make sure it is attached pr...

Page 6: ...e pump is due for service AlArms TroubleshooTing Continued TheCADD legacy Pluspumpsignalswithaudiblealarmsandmessages thatappearonthedisplaywhenaconditionrequiresyourattention upstream occlusion Two tone alarm Air in line Detected Two tone alarm stopped Three beeps every five minutes Programming incomplete Value not saved error Two tone alarm no message Displayed Two tone alarm no Disposable Clamp...