24
Alarms and Messages
If there are alarms or special messages you need to be aware of, the
pump will beep or sound an alarm. Look at the screen and follow the
steps in this table, or press
Å
for help.
When you see:
Take this action:
9 volt Battery
Low
NEXT to continue
The 9 volt battery is low but the
pump is operable. Change the
9 volt battery soon.
9 volt Battery
Depleted
Install good battery
Install a new 9 volt battery.
The pump will not run with a
depleted 9 volt battery. A good
battery must be installed even
when an external source of
power is connected.
(blank display or
random characters)
A two-tone alarm is sounding
and the amber light stays on.
The 9 volt battery is depleted.
Install a new battery.
Power Pack
Depleted
Change Power Source
The Power Pack is depleted.
Recharge the Power Pack.
Cassette Damaged
Free flow may occur
Clamp Tubing
NEXT to silence
The pump detects the cassette
is damaged. Close the tubing
clamp and inspect the cassette
for damage. Replace it if
necessary.
40-3665-01G PIG, Prizm TPN,VIP.indd 24
6/14/2011 10:50:29 AM
Summary of Contents for 6100