17
Attaching a Cassette
WARNING:
• If you are using a CADD
®
Administration Set or CADD™
Medication Cassette Reservoir that does not have the flow
stop feature (reorder number does not start with 21‑73xx):
You must use a CADD
®
Extension Set with Anti-Siphon Valve
or a CADD
®
Administration Set with either an integral or add
on Anti-Siphon Valve to protect against unregulated gravity
infusion that can result from an improperly attached cassette.
Unregulated gravity infusion can result in death or serious
injury.
• For detailed instructions and warnings pertaining to
the CADD™ Medication Cassette Reservoir or CADD
®
Administration Set, please refer to the instructions for use
supplied with those products.
CAUTION: Frozen medication must be thawed at room
temperature only. Do not heat the CADD™ Medication Cassette
Reservoir in a microwave oven as this may damage the
medication, the CADD™ Medication Cassette Reservoir, or cause
leakage.
Use aseptic technique as instructed by your clinician.
1. Clamp the tubing on the
new CADD™ Medication
Cassette Reservoir or CADD
®
Administration Set. If required,
remove blue clip from flow
stop feature.
2000-03-08 D. Zurn
«Prizm Hinge Cass»
40-3665-01G PIG, Prizm TPN,VIP.indd 17
6/14/2011 10:50:27 AM