25
When you see:
Take this action:
Air in line detected
Pump will not run
NEXT to silence
There is air in the tubing or the
tubing is not threaded through
the air detector. Press
„
to stop the alarm. Then follow
your clinician’s instructions for
priming.
Delivery Stopped
(Model 6101 only)
Fluid is not flowing from the
fluid container to the pump.
Check for a kink, a closed clamp,
or an air bubble in the tubing
between the fluid container
and pump. Press
⁄
to stop
the pump and silence the alarm
for 2 minutes, then remove the
obstruction and press
„
to
restart the pump.
High Pressure
Pump Stopped
There may be a kink in the
tubing or a clamp may be
closed. Unkink the tubing or
open the clamp and the pump
will resume delivery. You may
press
⁄
to stop the pump
and silence the alarm for
2 minutes. After you remove
the cause of high pressure, start
the pump if necessary. If the
alarm continues, contact your
clinician.
(any screen)
You hear 9 beeps: the
infusion profile has ended.
Disconnect or change the
CADD
®
Administration set as
appropriate.
Alarms and Messages