Accessibility of documentation
Please ensure that all instructions are kept in an easily accessible place for
operating personnel.
The operator checks and decides
All the information provided by the TRIGEN™ SURESHOT™ Distal Targeting System is to
help the operator make decisions during the operation. The operator must check all
the suggestions made by the system and is responsible for all decisions taken.
Responsibility of Smith & Nephew Orthopaedics
In the event of improper use, Smith & Nephew accepts no responsibility or liability
whatsoever for the functioning or utility of the TRIGEN SURESHOT Distal Targeting
System when used in the operating theatre.
Cleaning and sterilization
All instruments must be sterilized before use. Detailed information on the cleaning
and sterilization of components is contained in the separate
Cleaning and
Sterilization Instructions
(Smith & Nephew document 7138-1339).
Repair or modifications to the system
The user is not permitted to modify or service the equipment. There are no
serviceable parts inside the unit. Refer all service to authorized personnel.
Modifications/additions to the software
The user is not permitted to install or uninstall software. Any new software must be
installed by the manufacturer or by authorized personnel.
It is only allowed to connect equipment to the interface and power supply
connections of the TRIGEN SURESHOT Distal Targeting System which are IEC60601-1
approved and which are approved by Smith & Nephew Orthopaedics. Do not
modify this equipment without authorization of the manufacturer.
Electrical safety warning
To avoid risk of electric shock, this equipment must only be connected to a supply
mains with protective earth (=ground).
Avoid spilling water or other liquids on electronic/electrical equipment.
Use only Smith & Nephew disposables and accessories with the Smith & Nephew
TRIGEN SURESHOT Distal Targeting System
Maintenance
To verify accurate functionality, the device should be checked per the Maintenance
section of this document (beginning on page 19). This accuracy check must be
performed at least once every 12 months.
If this accuracy check is not performed as defined in the previous paragraph, all
warranty claims expire and the device is operated at the user’s own risk.
Recycling
Old electrical and electronic equipment must be disposed separately and may not
be included in the regular domestic waste. Alternatively, the unit can be handed
over to Smith & Nephew Orthopaedics for correct recycling.