14
Surgical procedure – Distal Targeting
Probe and set stop insertion
Following the removal of the Ball Tip Guide Rod,
insert the probe and set stop into the IM nail
as described in the
locking hole accuracy
check on the previous page.
Skin incision
Position the serrated tip of the drill sleeve
(represented by the green circle) over the desired
locking hole and make the initial incision.
Targeting the locking hole
With the appropriate length
TRIGEN
SURESHOT™ 4.0mm Drill Bit (7169-2810 or
7169-2811)
or the
TRIGEN SURESHOT Humeral
AO Drill Bit (7169-1155) inserted into the Targeter,
insert the tip of the drill sleeve (represented by
the green circle) through the incision and down
to bone.
Note When the Targeter is out of the preferred range or there is metal or electrical interference, the
green and red Targeter circles on the navigation screen may become unstable and/or a warning
message will be displayed. If the interference is excessive, the IM nail image on the navigation screen
will disappear. If interference cannot be avoided, a standard X-Ray technique must be used.
Note All tool cables should be uncoiled completely and any excess cables should be kept out of the
Targeter measurement volume.
Note Reference the warnings and tips in
the locking hole accuracy check on the
previous page.
Tip If performing this procedure using any of the
TRIGEN™ femoral antegrade nails (TAN™/FAN/
Pediatric/Adolescent), the Anteversion Locking
Guide (7169-2816) should be used along with a
4.0mm Drill Sleeve/Drill Bit to keep the IM nail
from rotating.
Critical Verify there are no other metal
objects (including metal triangles) in the field.
Metal interference will cause the system to
be inaccurate.
Summary of Contents for TRIGEN SURESHOT
Page 1: ...User Manual V2 1...
Page 2: ......
Page 37: ...35 Notes...
Page 38: ...36 Notes...
Page 39: ......