20
4
A software window will appear informing
the user if the TRIGEN™ SURESHOT™ Targeter and
Probe combination is within the
predefined accuracy parameters (“Pass”
or “Fail” message).
5
If the field accuracy check fails, check the
“Troubleshooting” section of this document for
possible solutions.
Software upgrades
Software upgrades are performed using a
Smith & Nephew, Inc. supplied USB memory
stick and a Work Instruction document. The Work
Instruction document should be filled out and
returned to Smith & Nephew Customer Service
upon completion.
Cleaning and disinfection
The TRIGEN SURESHOT Trauma Interface is used
in the non-sterile area of the operating theatre
and is cleaned and disinfected using commercial
cleaning and disinfectant products such as a mild
detergent and water or a bactericidal cleaning
solution such as 70% isopropyl alcohol.
Care must be taken not to allow any liquid to
pass into any electrical connections or the
interior of the unit. Let the surfaces dry
thoroughly before plugging in the unit. DO NOT
steam sterilize the system. DO NOT submerge
the system for any reason.
All other reusable instruments shall be cleaned
and sterilized according to the
Cleaning and
Sterilization Instructions
(Smith & Nephew
document 7138-1339).
Note Observe the manufacturer’s instructions for
dilution, exposure time, etc.
Note DO NOT sterilize with ethylene oxide gas or
steam sterilize the Trauma Interface.
Note Probes are NOT reusable.