User Manual
GW2145 - GW1160 - GW4060 - GW4190
Page 70 - 86
10
TECHNICAL DATA
The documentation includes a description of the rated characteristics of the electrical connection, all the intake and
outlet connections and the ambient conditions for which the appliance was designed (ref. point 5.4.2 of IEC61010-
1:2001). For “non electric” intakes, the pressure and flow ranges are provided (ref. point 5.4.2 IEC61010-2-040:2005).
WATER SUPPLY REQUIRED
Type of water required
1 - Mains (required)
2 -
Demineralised
(recommended)
3 - hot
Pressure
200 kPa – 500 kPa (2 bar – 5 bar)
Type of threaded connection
3 / 4 ”
Required flow [min - max]
2 - 12 litres / min
Mains water hardness [max]
(for models equipped with a steam condenser, the
use of purified water with hardness below 15°f is
recommended).
42 °f
-
42 °f
Fe2+ / Fe3+ [max]
0.5 ppm
water temperature [max]
35°C
35°C
35°C
Demineralised water conductivity [max]
-
30μS/cm
-
Limits depending on microbial
contamination
Minimum microbiological quality required: “potable water" (ref. limits
in accordance with Italian Legislative Decree 31/2001)
DIMENSIONS and WEIGHT
Series
GW2145
GW1160, GW4060
GW4190
Standard External
[Height x Width x
Depth]
850 x 450 x 620
850 x 600 x 600
850 x 900 x 600
Standard External - aquastop accessory
[Height x Width x Depth] - overall height is
increased by
+7mm
-
857
x 600 x 600
857
x 900 x 600
Standard External - PAD1 accessory
[Height x Width x Depth] - overall depth is
increased by
+110mm / +150mm
(depending on the model of PAD
installed).
850 x 450 x
730 / 770
850 x 600 x
710 / 760
850 x 900 x
710 /
760
Standard External - peristaltic pump P3
accessory
[Height x Width x Depth] – Only
for GW1160, GW4060: overall depth is
increased by
+60mm
-
850
x 600 x
660
850 x 900 x 600
Net weight
56 kg
72 kg
105 kg
Weight of app packaging
66 kg
90 kg
120 kg
MATERIALS
Washing chamber
AISI 316L
External surface
AISI 304