![Smeg B-One User Manual Download Page 43](http://html1.mh-extra.com/html/smeg/b-one/b-one_user-manual_1303010043.webp)
Revision n.1
User Manual B-One
Edition: 04/2009
Page 43 of 52
THIRD
GROUP
DESCRIPTION
6
Door opening phase: decompression.
Error may have been generated by the presence of an object located
between the door and lining. In this case, remove object and check that
front lining has not been damaged.
7
Door opening phase: recovery of atmospheric pressure.
Error may have been generated for banal reasons, such as obstruction or
improper positioning of the bacteriological filter (the filter is the round, white
component located under the openable door, can be unscrewed by hand
for removal)
8
Door opening phase: motor movement, moving away from alignment.
9
Door opening phase: motor movement, closer to alignment.
10
Door closing phase: motor movement, moving away from alignment.
11
Door closing phase: motor movement, closer to alignment.
12
Water load phase into loading tank.
The minimum level of water has not been reached in the expected time.
Check to see if the demineralised water can is empty and check that the
withdrawal tube end is still immersed in the water. Check that loading tube
is not blocked.
13
Water draining phase from discharge tank. Check that discharge tube is
not blocked.
14
Water load phase into steam generator.
15
Vacuum application phase for B&D/helix test cycles and for vacuum tests.
16
Vacuum maintenance phase in the vacuum test cycle.
If the test has been performed with a warm machine, we recommend
proceeding with another test after the machine has been cooled down.
Vacuum tests are best performed immediately after machines have been
started up under cooler conditions.
18
Error during fractioned vacuums - initial cycle phase.
Error may have been generated by the presence of an object located
between the door and lining. In this case, remove object and check that
front lining has not been damaged.
19
Pressure recovery phase during fractioned vacuums.
Error may involve function of the machine solenoid valve. Keep in mind that