5.2
TURN-ON
The autoclave switches on when the ON/OFF button (blue general switch) is pressed. The
SMEG logo appears on the display and a BEEP sound will be activated.
The autoclave carries out a routine initial check. Any irregularities will be shown on the
display.
fig. 9 – Start-up screen – Smeg logo
Immediately following the initial check, the preheating phase starts. The main cycle
selection screen will appear on the display (see the following "preheating" paragraph for
further details).
fig. 10 – Main screen – cycle selection
It is also possible to select and confirm sterilization cycles during the preheating phase.
The machine will automatically manage the initial wait (approx. 15 min) and cycle start-up.
The vacuum test cycle does not need to wait until the end of the preheating phase to be
turned on.
5.3
STERILIZATION CYCLE SELECTION AND START-UP
Select the desired cycle from the main screen by turning the knob. Press the knob to
confirm cycle.
fig. 11 – Main screen
Once the desired cycle has been selected, the interface will request confirmation to
proceed.