-56-
No.LFE****-OMZ0013
<When unlocking the key lock>
* The following procedure of cancelling the key lock should be followed when [F81] Security code request is
set to valid.
(1) Press the
button for 5 seconds or longer in measurement mode.
The current setting [LoC] flashes on the sub display.
Sub screen
(2) Press the
or
button to select unlocking [UNLoC].
(3) After the
button is pressed, the security code must be entered.
(4) Input of security code (3 digit setting)
The first digit will start flashing.
Press the
or
button to change the value.
Press the
button to make the next value to the right flash.
(If the
button is pressed on the far right digit, the hundreds digit
will flash)
(5) After the input is complete, press and hold the
button for 1
second or longer. The security code will be confirmed.
(If no key operation is performed for 30 seconds during input or
change of the security code, the display will return to
measurement mode with LoC status.)
If the security code entered is wrong, [FAL] will be indicated on the
sub display.
In this case, retry inputting the security code.
If an incorrect security code is entered 3 times, the display will
return to measurement mode with LoC status.
Enter 000 for the default setting.
[UNLoC] is displayed on the sub screen.
(6) Press the
button to complete the unlocking operation. Return to measurement mode.
Locking Unlocking
Summary of Contents for LFE0 Series
Page 1: ...Doc No LFE OMZ0013 Product Name Flow Monitor Model Series Product Number LFE0 SMC Corporation ...
Page 36: ... 35 No LFE OMZ0013 List of output modes for fluid temperature ...
Page 42: ... 41 No LFE OMZ0013 F20 External input This item is not used for this specification ...
Page 66: ... 65 No LFE OMZ0013 Dimensions Front protective cover Panel mount adapter R ...