-16-
No.SFOD-OML0002-A
○
Error indication function
Error Name
Error display
Error Type
Troubleshooting Method
Zero clear
error
A charge over ±5%F.S. of the default
zero point is given to the sensor during
zero clear operation. (Er1)
∗
: After approx. 1 s. error display, the sensor
returns to measurement mode.
This time will depend on the deviation of
the product itself and ambient environment
during zero clear operation.
Return to a condition without charge,
and retry zero clear operation.
Sensor failure
The sensor is broken.
Stop using immediately and contact
SMC.
System error
Error in internal data.
Turn off the power supply once, and
turn it on again.
Ask SMC to investigate if the error
persists.
The value is outside of the displayed
range because a charge over the upper
limit of the measurable voltage range is
given to the sensor or the measuring
distance is not appropriate.
Measurement
error
The value is outside of the displayed
range because a charge over the lower
limit of the measurable voltage range is
given to the sensor or the measuring
distance is not appropriate.
Eliminate charge until the value
comes to the measurable voltage
range. Also, check the measuring
distance is proper.
Cable broken
The cable is broken and the product
cannot measure correctly. The charge
amount is detected, but not reflected in
the displayed value.
Stop using immediately and contact
SMC.