Specifi cations: 

Air-cooled Refrigeration

Model

HRR012-A

HRR018-A

HRR024-A

HRR030-A

Cooling method

Air-cooled refrigeration

Refrigerant

R410A (HFC)

Refrigerant charge

[kg]

0.38

0.47

Control method

PID control

Ambient temperature/Humidity/Altitude

1, 13

Temperature: 5 to 40 

°

C, Humidity: 30 to 70 %, Altitude: less than 3000 m

Cir

culating fl

 uid 

system

Circulating fl uid

2

Tap water, 15 % ethylene glycol aqueous solution

Set temperature range

1

[

°

C]

10 to 35

Cooling capacity (50/60 Hz)

3

[W]

1000/1200

1600/1800

2000/2400

2500/3000

Heating capacity (50/60 Hz)

4

[W]

450/500

550/700

Temperature stability

5

[

°

C]

±

0.1

Pump capacity (50/60 Hz)

6

[MPa]

0.13 (at 7 l/min)/0.18 (at 7 l/min)

For Option T: 0.42 (at 10 l/min)/0.4 (at 14 l/min)

For Option MT: 0.32 (at 10 l/min)/0.32 (at 14 l/min)

Rated fl ow (50/60 Hz)

7

[l/min]

7/7  For Options T, MT: 10/14

Flow display range

[l/min]

3 to 16

Electric conductivity display range

[

μ

S/cm]

2 to 48 (Only for Option DM)

Electric conductivity setting range

[

μ

S/cm]

5 to 45 (Only for Option DM)

Particle fi lter nominal fi ltration rating [

μ

m]

5

Bypass valve

Installed

Tank capacity

[L]

Approx. 4

Outlet, Return port size

Rc1/2

Drain port size

Rc1/4, With cap

Leakage protection

Drain pan (With water leakage sensor)

Fluid contact material

Stainless steel, Copper (Heat exchanger brazing)

11

, Brass

16

, SiC, 

Alumina ceramic, Carbon, PP, PE, POM, PA, FKM, EPDM, PVC, 

PPS, AS, Fluoropolymer

12

, Ion exchange resin

12

Electrical system

Power supply

Single-phase 200 to 230 VAC, 50/60 Hz, Allowable voltage range 

±

10 %

14

Circuit protector

[A]

10  For Options T, MT: 15

Applicable earth leakage breaker capacity

8

Rated current: 10 A  For Options T, MT: 15 A

Sensitivity current: 30 mA

Cable quantity x Size (Including grounding cable)

15

3 cores x 14 AWG (3 cores x 2.0 mm

2

)

Rated operating current
(50/60 Hz)

3

[A]

3.6/4.0

3.6/4.3

5.2/5.8

5.5/6.2

For Options T, MT

4.7/6.2

4.7/6.5

6.3/8.0

6.5/8.3

Rated power consumption
(50/60 Hz)

17

[kW(kVA)]

0.6/0.8 (0.7/0.8)

0.7/0.9 (0.7/0.9)

0.9/1.2 (1.1/1.2)

1.0/1.2 (1.1/1.3)

For Options T, MT

0.9/1.2 (0.9/1.2)

0.9/1.3 (1.0/1.3)

1.2/1.6 (1.3/1.6)

1.3/1.6 (1.3/1.7)

Communication function

Contact input/output, Serial RS-485/RS-232C

Noise level (50/60 Hz)

9

[dB]

59/60

59/60

61/64

61/64

Accessories

18

Power supply connector, Particle fi lter element and maintenance handle, Operation manual

Weight

10

[kg]

40

46

1   It should have no condensation. During seasons or in locations where 

the ambient temperature is likely to fall below freezing point, please con-
tact SMC for that case.

2   If tap water is used, use water that conforms to Water Quality Standards 

of the Japan Refrigeration and Air Conditioning Industry Association 
(JRA GL-02-1994 cooling water system - circulating type - make-up wa-
ter).

3   

q

 Ambient temperature: 2 5 

°

C, 

w

 Circulating fl uid temperature: 2 0 

°

C, 

e

 Circulating fl uid rated fl ow, 

r

 Circulating fl uid: Tap water, 

t

 Power 

supply: 200 VAC, 

y

 Piping length: Shortest

For models with high-pressure pump mounted (Options T, MT), the cool-
ing capacity will decrease by 300 W.

4   

q

 Ambient temperature: 2 5 

°

C, 

w

 Circulating fl uid temperature: 2 0 

°

C, 

Circulating fl uid rated fl ow, 

r

 Circulating fl uid: Tap water, 

t

 Power 

supply: 200 VAC, 

y

 Piping length: Shortest

5   Temperature at the thermo-chiller outlet when the circulating fl uid fl ow is 

rated flow, and the circulating fluid outlet and return port are directly 
connected. Installation environment and the power supply are within 
specifi cation range and stable.

6   The capacity at the thermo-chiller outlet when the circulating fl uid tem-

perature is 20 

°

C.

7   Required fl ow rate for maintaining the cooling capacity or temperature 

stability
The specifi cation of the cooling capacity and the temperature stability 
may not be satisfi ed if the fl ow rate is lower than the rated fl ow.

8   Purchase an earth leakage breaker with sensitivity current of 30 mA/pow-

er supply 200 VAC separately.

9   Front: 1 m, height: 1 m, stable with no load, Other conditions 

J

 See 

3.

10   Weight in the dry state without circulating fl uids

The weight will increase by 1 kg when Option DM (With electric con-
ductivity control fu Applicable to DI water piping) is selected.
The weight will increase by 5 kg when Option T (High-pressure pump 
mounted) or MT (Applicable to DI water High-pressure pump 
mounted) is selected.

11   Copper and brass are not included when Option M (Applicable to DI 

water piping) or MT (Applicable to DI water High-pressure 
pump mounted) is selected.

12   For Option DM (With electric conductivity control fu Applicable 

to DI water piping)

13   If the product is used at altitude of 1000 m or higher, refer to “

  For al-

titude of 1000 m or higher” on page 25.

14  No continuous voltage fl uctuation

15  To be prepared by the customer

16  For Option T (High-pressure pump mounted), this material is included.

17   

q

 Ambient temperature: 2 5 

°

C, 

w

 Circulating fl uid temperature: 2 0 

°

C, 

e

 

Circulating fl uid rated fl ow, 

r

 Circulating fl uid: Tap water, 

t

 Power supply: 

200 VAC, 

y

 Piping length: Shortest, 

u

 At the rated cooling load applied

18   For Option DM (With electric conductivity control fu Applicable 

to DI water piping), a DI fi lter is attached.
For pipe thread type F, a G thread conversion fi tting set is attached.
For pipe thread type N, an NPT thread conversion fitting set is at-
tached.

10

Thermo-chiller/Rack Mount Type  

HRR 

Series

Summary of Contents for HRR Series

Page 1: ...bility Single phase 200 to 230 VAC 50 60 Hz Global power supply 1 0 1 6 2 0 2 5 kW 50 Hz Air cooled Refrigeration Cooling capacity DI filter Option Particle filter Standard Bypass valve Standard 310 m...

Page 2: ...duct Optional Accessories Optional Accessories Cap Drain port Rc1 4 Water leakage sensor Drain pan Handle Accessory Replacement particle filter element DI filter replacement Drain pan Built in water l...

Page 3: ...Set value displayed in green Set temperature etc is displayed Easy check from the circulation fluid level HRS012 HRR012 HRR012 HRS012 615 mm 310 mm Flow rate and pressure adjustment can be seen on th...

Page 4: ...opens and closes to keep the refrigerant gas pressure consistent The facility water flow rate is controlled by the water regulating valve Facility water circuit Circulating fluid circuit Refrigeration...

Page 5: ...LEVEL a WRN Alternately displayed Maintenance notification number MT01 X is displayed if a maintenance notification has occurred To menu screen REF a TEMP Alternately displayed Compressor inlet temper...

Page 6: ...d can be output Remote operation signal input Chiller can be operated remotely by contact input signals Output setting example Output 1 Temperature rise Output 2 Pressure rise Output 3 Operation statu...

Page 7: ...wave inspection machine Transmission cable connector for fibre laser Cooling of the laser irradiated part Temperature control of the X ray tube and X ray light sensing part Cooling of the UV lamp X r...

Page 8: ...on Water cooled Refrigeration HRR Series Pressure Switch Monitors pressure of the circulating fluid and facility water ISE20C ISE7 G 3 Screen Display High Precision Digital Pressure Switch PSE300AC PS...

Page 9: ...Alarm p 16 Communication Functions p 17 Thermo chiller Rack Mount Type HRR Series Options With Electric Conductivity Control Function Applicable to DI Water Piping p 19 Applicable to DI Water Piping...

Page 10: ...pressure pump mounted U Conforming to UL Standards When multiple options are combined indicate symbols in alphabetical order Water cooledRefrigeration HRR U W 20 Pipe thread type Rc F G With Rc G conv...

Page 11: ...of the Japan Refrigeration and Air Conditioning Industry Association JRA GL 02 1994 cooling water system circulating type make up wa ter 3 q Ambient temperature 2 5 C w Circulating fluid temperature 2...

Page 12: ...e 2 If tap water is used use water that conforms to Water Quality Standards of the Japan Refrigeration and Air Conditioning Industry Association JRA GL 02 1994 cooling water system circulating type ma...

Page 13: ...1500 1000 500 0 Ambient temperature or facility water temperature 32 C 25 C 40 C Circulating fluid temperature C Cooling capacity W 0 10 20 30 40 3500 3000 2500 2000 1500 1000 500 0 Ambient temperatur...

Page 14: ...emperature or facility water temperature 40 C 32 C 25 C 5 C Circulating fluid flow rate l min Circulating fluid pressure MPa 0 5 10 15 20 25 0 3 0 2 0 1 0 Outlet 60 Hz Outlet 50 Hz Return port Allowab...

Page 15: ...connector Product label Circulating fluid outlet Rc1 2 226 73 424 397 Ventilation air outlet Ventilation air inlet 361 4 259 8 139 1 37 5 0 86 10 440 Includes protruding parts 550 50 Maintenance port...

Page 16: ...r Inside of the cover panel 210 38 0 424 397 164 5 140 5 0 272 5 183 6 126 4 37 5 0 440 Includes protruding parts 550 24 50 Maintenance port Power supply switch Operation display Fluid level gauge 10...

Page 17: ...9 key Move item downward or decrease the set value 0 RESET key Reset the alarm Alarm code Alarm message Initial value Display unit Upper stage White Lower stage Green AL01 Low level in tank abnormal F...

Page 18: ...pplication Division Initial value Default setting 1 24 VDC output Output 2 24 VDC output Output 3 24 VDC output Output 4 Contact input signal 1 Input OFF 5 Common of contact output signal 1 Output 6 C...

Page 19: ...D 3 SG 5 This product RD 2 SD 3 SG 5 D sub 9 pin female socket type Communication connector 5 1 9 6 Run Stop Circulating fluid temperature setting Writing Circulating fluid discharge temperature Circu...

Page 20: ...ng fluid circuit is made from non copper materials Available DI water is electric conductivity 2 S cm or more Electric resistivity 0 5 M cm or less This Option M does not have electric resistance elec...

Page 21: ...p Capacity HRR012 018 024 030 20 MT HRR012 018 024 030 20 T Applicable model HRR012 018 024 030 20 T HRR012 018 024 030 20 MT 1 Pump Rated flow 50 60 Hz 2 3 l min 10 0 42 MPa 14 0 40 MPa 10 0 32 MPa 1...

Page 22: ...rtridge Element for the maintenance of the particle filter for circulating fluid DI filter cartridge for replacement for Option DM Electric conductivity control type DI water piping type Part number A...

Page 23: ...ser s equipment Q qm x C x T2 T1 x qv x C x T 60 1 x 10 x 4 2 x 103 x 2 0 60 1400 J s 1400 W Cooling capacity Considering a safety factor of 20 1400 W x 1 2 1680 W Heat generation amount by user s equ...

Page 24: ...the circulating fluid discharge pressure Consider the installation height difference between the thermo chiller and the user s equipment and the piping resistance such as circulating fluid pipings or...

Page 25: ...arts such as piping may cause clogging or leakage in the circulating fluid circuit Therefore take sufficient care when selecting fluid contact part materials such as piping 4 Design the piping so that...

Page 26: ...he descriptions in the ta ble below Select the thermo chiller considering the descriptions qUpper limit of ambient temperature Use the product in ambient temperature of the described value or low er a...

Page 27: ...be connected to a water line gas line or lightning rod HRR Series Specific Product Precautions 3 Mounting Installation Warning 1 Do not use the product outdoors 2 Do not place heavy objects on top of...

Page 28: ...iping due to metal ion 5 Do not use fluid that includes metric powers and other foreign materials It can cause issues clogging of the circulating fluid circuit or leakage Warning HRR Series Specific P...

Page 29: ...er once again when the fluid level is reduced When there is no reduction in the fluid level the job of removing the air is completed 2 Confirmation during operation Check the circulating fluid tempera...

Page 30: ......

Page 31: ...nergy railways air navigation space shipping vehicles military medical treatment combustion and recreation or equipment in contact with food and beverages emergency stop circuits clutch and brake circ...

Reviews: