Programming Using Local adjustment
4.9
RANGE BRANCH (RANGE)
Z:
Moves to the FUNCT function.
S:
Enters the RANGE branch, starting with the function UNIT.
Unit (UNIT)
Z:
Moves to the LRV function.
S:
Starts selection of engineering unit for process variable and setpoint indication. After
activating (
S
), you can move around the options available in the table below by
activating (
Z
). Using (S) activates the desired unit. Escape leaves the unit unchanged.
UNIT
DISPLAY
DESCRIPTION
InH
2
O
Inches water column at 20
o
C
InHg
Inches mercury column at 0
o
C
ftH
2
O
Feet water column at 20
o
C
mmH
2
O
millimeter water column at 20
o
C
mmHg
millimeter mercury column at 0
o
C
psi
pounds per square centimeter
Bar
Bar
Mbar
millibar
g/cm
2
grams per square centimeter
k/cm
2
Kilograms per square centimeter
Pa
Pascals
kPa
Kilo Pascals
Torr
Torr at 0
o
C
atm
atmospheres
ESC
-escape-
Table 4.4
– Units
Lower Range Value Adjustment without Reference (LRV)
Z:
Moves to the LRV DECREASE function.
S:
Increases the Lower Value until the magnetic tool is removed or the maximum for the
Lower Value is reached.
Z:
Moves to the URV ADJUSTMENT function.
S:
Decreases the Lower Value until the magnetic tool is removed or the minimum for
the Lower Value is reached.
Summary of Contents for ld301
Page 14: ...LD301 Operation and Maintenance Instruction Manual 1 8...
Page 18: ...LD301 manual de Instru es Opera o e manuten o 2 4 Fig 2 3 LD301 Software Block Diagram...
Page 50: ...LD301 Operation and Maintenance Instruction Manual 4 16...
Page 55: ...Maintenance Procedure 5 5 Fig 5 1 Exploded View...
Page 62: ...LD301 Operation and Maintenance Instruction Manual 5 12...
Page 70: ...APPENDIX A...