SlimLine Force Sensor, Type 9173B to 9177B
Page 28
9173B_002-286e-07.10
7.2
Repairing SlimLine Force Sensor
The SlimLine force link
cannot
be repaired if the following
behavior is being exhibited:
No output signal present
Wrong measurement due to reduced sensitivity (indica-
tion of malfunction caused by overload)
Pronounced drift that cannot be eliminated by cleaning
the connector
Please note that repairs of certain types of damage that
cannot be undertaken cost-effectively will not be carried
out by the manufacturer's factory.
Fig. 13: Cable connection
Fig. 14: Cable break
broken
off
If your SlimLine sensor is faulty and is not damaged as
shown in Figures 13 und 14 please adopt the following
procedure:
Contact your Kistler distributor to notify return of the
faulty instrument.
Return the faulty sensor to the distributor in its original
packaging.
Enclose a detailed description of the fault and the at-
tendant circumstances with the sensor.
Describe the measurement operation during which the
fault arose.
You will receive a cost estimate if a major repair is in-
volved.
Kistler will try to repair your sensor at minimal cost in
the shortest possible time and return it to you in as-new
condition.