SlimLine Force Sensor, Type 9173B to 9177B
Page 26
9173B_002-286e-07.10
6.2
Kistler Calibration Service
To maintain their position on the world market products
have to meet extremely stringent quality requirements. ISO
9001 requires test equipment to stand in a known and
valid relationship to nationally recognized standards.
Traceability must be ensured. Kistler offers the following
calibration services:
Swiss Calibration Service (SCS)
Kistler is accredited as SCS Calibration Center No. 049 for
equipment measuring pressure, force, acceleration and
electrical charge. To ensure traceability and specified
measurement uncertainties, calibration equipment and
methods are regularly monitored and audited.
In-Situ Calibration
Kistler offers an in-situ calibration service for cases in which
the instruments and equipment to be calibrated cannot be
transported.
Calibration in One of the Calibration Centers
In addition to in-situ calibration Kistler offers a service that
allows customers to have transportable instruments and
equipment calibrated in one of the calibration centers.
6.3
Maintenance and Service
Please observe the following instructions for maintaining
your SlimLine quartz force link
Check the SlimLine force link each time before use for
obvious damage. Do not operate a damaged sensor!
Leave all repairs to the force link (such as replacing
damaged connectors) to the manufacturer's factory.
You will find details of the procedure to be adopted in
such a case in Section 7.2.
The bottom and top surfaces of the SlimLine force links
are ground parallel and flat to within very tight toler-
ances. These surfaces must not be refinished! Visible
scratches and grooves can be leveled out with a grind-
stone.
Clean the connector of the SlimLine force link or the
connecting cable regularly with a completely unadulter-
ated cleaning agent, such as the cleaning spray Type
1003 from Kistler or white spirit.
Summary of Contents for 9173B
Page 1: ...Instruction Manual SlimLine Force Sensor Type 9173B to 9177B 9173B_002 286e 07 10...
Page 2: ...Instruction Manual SlimLine Force Sensor Type 9173B to 9177B 9173B_002 286e 07 10...
Page 27: ...Calibration and Maintenance 9173B_002 286e 07 10 Page 25 Fig 12 In situ calibration procedure...