17. SLE2000 Trouble Shooting Chart
Symptom
Possible Cause
Remedy
Leak Alarm - audible
& visual
Incorrect circuit fitted.
Damaged circuit fitted.
Humidifier chamber.
High alarm condition.
Fresh gas supply has
decreased.
Leak alarm drift
Replace with correct circuit
P/No. N2188 or variant.
Replace with new circuit.
Replace chamber.
High alarm LED illuminated
Reset Insp. pressure and/or
adjust High alarm parameter
Consult Service Manual.
Consult Service Manual.
Block Alarm -
audible & visual
Fresh gas filter blocked.
Fresh gas restrictor blocked.
Replace filter.
Replace circuit.
Blender Alarm -
audible when both
gases connected.
Alarm shuttle sticking.
Either Air or O
2
gas has failed
Air/O
2
differential pressure
exceeded.
Remove both gases from supply
and re-connect O
2
first. If this
does not work, disconnect and
then connect the Air first.
Consult SLE trained Engineer if
alarm persists.
Repair gas supply
Check that the difference in
pressure between the two
gasses is less then 30 PSI.
High Alarm - audible
& visual
Incorrect PIP alarm setting.
Readjust alarm to INSP
PRESSURE or readjust INSP
Pressure regulator.