8.5 CMV Mode
Set the following ventilator conditions :-
CPAP/PEEP
regulator to 10 cmH
2
O
Re-set
Inspiratory
regulator to
30
cmH
2
O
Set the Pressure display switch to
PEAK
Set
PIP alarm control
to 40 cm H
2
O
Advance the mode switch to
CMV
Set the
BPM
Rate to 60
Set the
INSP TIME
to 0.50
Ensure that :-
•
The ventilator is cycling
•
I:E Ratio display reading 1:1
•
With the pressure switch check that the mean reads approximately 20 and the peak
reads approximately 40 on the LED.
8.6 PTV Mode
Set
BPM
rate to
20
in
CMV
mode
Advance the mode switch to
PTV
Set the
PTV Sensitivity
control to
1
Ensure that :-
•
Ventilator should continue to cycle
• BPM
display reading is initially 0
• TRIGGER BACK-UP LED
is “on”