22
CONSUMABLES
IOM cassette, plastic
225-71A
Foam inserts for Respirable selection
225-772
Other selector foams not currently available ---
GFA 25 mm filters
225-58F
PARTS & ACCESSORIES
Sidekick air sampling pump
224-50
Sidekick air sampling pump with timer
224-52TX
(224-52TX is an intrinsically safe pump)
Charger for Sidekick
223-203C*
Tygon flexible tubing ( one metre)
225-13-4A
DC Lite calibrator
717-01K
Rotameters also available - see options in the latest catalogue
Calidaptor to connect DC Lite/rotameter
390-01
Replacement SPLIT2 Battery Pack
770-303
SPLIT2 Battery Charger
770-310C*
Replacement data cable
770-114
Replacement calibration standard
770-207
Replacement zero filter
770-112
Manual
770-300M
*C denotes a UK plug. For a European 2 pin plug use B
This section is IMPORTANT as it covers the procedure involved to ensure the reliability and accuracy of your unit is optimised despite
the often unavoidable changes in temperature.
Drastic changes in the environment temperature could affect the readings from the Split2. Temperature changes can occur, for example,
when transferring between indoor and outdoor monitoring, or if the unit has been stored overnight in a cold room or vehicle, before
being used in a warm room.
Follow this procedure:
1.
Take the SPLIT2 to the area where you wish to sample
2.
Run the unit for 15-20 minutes with the zeroing filter on and the perform an auto zero.
3.
Remove zeroing filter.
4.
Choose overwrite option. This will erase all the previous memory locations already stored and allow the electronics to
stabilize (to the sampling temperature). The auto zero will activate the temperature compensation.
It is recommended to auto zero the unit each time the user changes location, this resets the temperature compensation to
ambient
CONSUMABLES & SPARES
TEMPERATURE CHANGES