14
Auto - Zeroing
Please Wait
24
Run
Review Data
Special Functions
Auto-Zero
The Split2 is now Auto - Zeroing. The time in seconds will appear
in the bottom right hand corner of the LCD indicating how long
for the Auto - Zero to be completed.
Once completed the SPLIT2 will return to the main menu
5. Setting the Alarm:
Go to the main menu screen. (As a starting point we recommend
the alarm is set at 80% of the Occupational Exposure Limit)
Run
Review Data
Special Functions
Auto-Zero
System Options
Date/Time
Set Alarm
Cancel
Scroll arrow to Set Alarm option and press the
ENTER
button
The ALARM is configured as concentration in Milligrams per cubic metre (Mg/m
3
)
To set all parameters on this screen the following procedure applies:
To increase the value of the selected digit... Press the
UP
button.
To decrease the value of the selected digit .... Press the
DOWN
button.
To confirm the digit and select the next field.... Press the
ENTER
button
When all the data is correct press
ENTER
The screen will now return to the Main Menu
Go to the Special Functions option by pressing either the down
button or the up button until the arrow is opposite Special
Functions. Press
ENTER
button
*
SET ALARM
*
Conc: 000.00 mg/m
3
Cursor
The following screen will appear with a cursor (line) under
the 0 character
Please NOTE: Alarm only works if the Alarm Continue
option is taken from the RUN menu. Please see section
on TAKING A SAMPLE page 17