6
Battery Operation
2. Carefully align the ba
tt
ery jack on
the replacement ba
tt
ery pack with
the ba
tt
ery terminal on the bo
tt
om
of the pump base plate and push
the ba
tt
ery pack into place.
3. Replace and tighten the two
security screws removed in
Step 1.
Use of a repaired or rebuilt battery pack voids the SKC warranty and the
UL Listing for intrinsic safety.
Do not charge or operate the pump from the charger in hazardous
locations!
Use only an SKC-approved charger and battery pack designed for the
AirChek 2000 Sample Pump to ensure reliable performance and to
maintain the UL Listing for intrinsic safety and the SKC warranty.
Use of any device other than the approved battery pack to power the
pump voids the UL Listing for intrinsic safety.
Battery Eliminator
The Ba
tt
ery Eliminator is an accessory that converts alternating current (AC)
to direct current (DC) from which the pump can operate inde
fi
nitely.
The
Ba
tt
ery Eliminator should be used in non-hazardous locations only.
See
page 28 for ordering.
To use the Ba
tt
ery Eliminator, the ba
tt
ery pack must be removed from the
pump
(see Replacing the Ba
tt
ery Pack on page 5
). The Ba
tt
ery Eliminator is
comprised of two pieces, (1) a wall cube that converts AC voltage to DC
voltage and (2) a power adapter that reduces DC voltage. The wall cube
fi
ts into a standard wall outlet and its plug end is inserted into the power
adapter. The power adapter is
fi
tt
ed on the pump in place of the ba
tt
ery
pack.
Use the battery eliminator to operate the pump in non-hazardous
locations only.
Use of any device other than the approved battery pack to power
the pump voids the UL Listing for intrinsic safety.
¡
For more information on battery care and maintenance, go to
http://www.skcinc.com/instructions/1756.pdf.
Ba
tt
ery
terminal
Pump base
plate
Ba
tt
ery
jack
Ba
tt
ery
pack
Summary of Contents for AirChek2000