22
Calibration
CalChek Error Chart
Single-point Calibration Errors
Error
Problem
Troubleshooting
E41
Correction required too large. A
Perform a full calibration. If this fails
mismatch of greater than 360 ml/min
contact SKC Technical Support.
between
the
fl
ow setting on the pump
and the reading generated by the
Defender Calibrator has occurred.
E48
Could not get a successful single point
Try the calibration again. If problem
calibration
within
fi
ve
fl
ow readings.
persists, perform a full calibration.
Full (Multiple-point) Calibration Errors
Error Problem
Troubleshooting
E44 First
fl
ow reading greater than 750
Check tubing between pump’s upper
ml/min. The pump is
fl
owing faster
pressure sensor and the top valve
than it should, even though the
assembly diaphragm to ensure it is not
calibration routine delivered only a
pinched or blocked, or contact SKC
very small voltage to the pump.
Technical Support.
E45
Pump unable to achieve
fl
ow rate of
Check pump’s
fl
ow tube to ensure
3000 ml/min possibly due to a blocked
it is not blocked, or contact SKC
fl
ow tube or an air leak inside the pump.
Technical Support.
E46
Analysis error in the data (rare).
Try full calibration again. If problem
persists, contact SKC Technical Support.
E47
Less than two bars appear in the battery
Recharge the battery.
icon on the pump display indicating that
the battery is too low. There must be at
least two bars to begin a full calibration.
No Code At conclusion of full calibration, pump
Pump not at ambient conditions for at
does not verify to within 5%.
least 2 hours. Retry calibration after
pump has been at ambient conditions
for 2 hours.
Pump not running for 2 minutes
prior to calibration. Run pump for
2 minutes and retry calibration.
Errors That Can Occur During Both Calibration Modes
Error Problem
Troubleshooting
E42
Unstable average. There is too
Try the calibration again. If problem
much variation in the
fl
ow readings.
persists, contact SKC Technical
Support.
E43
Serial time out. The calibrator is
Check cable connection. If loose
not communicating with the pump.
or disconnected, connect properly.
Summary of Contents for AirChek2000
Page 34: ...31 ...