11
ADJ
Used during calibration with primary
standard calibrator
(not for use with
CalChek feature
). Press
S
to increase
or
T
to decrease
fl
ow adjustment
until desired
fl
ow is indicated on an
a
tt
ached primary standard. Press
Ò
to
move to next parameter or the User Interface can be exited
by continuing to press
Ò
until End appears. Press [
ST
] to
save new
fl
ow and adjustment se
tt
ings and exit the User
Interface.
See page 16.
CAL
Use for CalChek calibration feature
only. Press [
ST
] to initiate single-
point calibration. Press
S
seven times
to start a full calibration.
See CalChek
Calibration instructions on pages 18 to 22
.
Using User Interface - Level Two
Temperature
Press
S
or
T
to toggle between
Fahrenheit (F) and Celsius (C). Press
Ò
to move to next parameter.
Atmospheric
Pressure
Press
S
or
T
to toggle through
inches of mercury (ins), millimeters
of mercury (mm), and millibars (m).
Press
Ò
to move to next parameter.
Sample Time
(
S
Г
)
Program a speci
fi
c run time. Press
S
to increase or
T
to decrease the
time in minutes (from 1 to 999). Press
Ò
to move to the next parameter.
Sample Time is not an option when a
DataTrac 2000 program is loaded into
pump memory (PROG).
See page 13.
12-hr/24-hr
Clock and
Press
S
or
T
to toggle through
Delayed Start
standard (12-hour) time, military
(24-hour) time, and delayed start
(dELA).
See Se
tt
ing a Delayed Start
on page 14.
Press
Ò
to move to
next parameter.
Setup