skcinc.com
Form 37165 Rev 200217
Page 5
Sampling Screen
Touch the PUMP key to begin sampling.
Touch the PUMP key again to pause or stop sampling.
Sampling Screen Indicators
Note
:
The
relative flow indicator is displayed in place of seconds on sampling screen when up or down arrow key is
touched during sampling. Also displayed on set/calibrate flow screen. See Set/Calibrate Flow Screen for details.
See
Options/Modes during Sampling for details about the screen lock feature and flow fault mode.
Reading Pump Status Indicators
The status LEDs that bracket the screen display (
see Figure 1
) indicate pump status:
Green, flashing
= Running
Red, flashing
= Flow fault
Note
: Status LEDs will flash red/green to indicate that the pump is out of flow tolerance just before entering flow fault
mode and during each auto-restart attempt while in flow fault mode.
Elapsed run time HH:MM:SS
Battery status
See Note
Screen lock
feature activated
Flow Fault or Low Battery Fault
Relative Flow
—
A number between 1 and 99 that is relative to a flow rate and changes as flow is
adjusted: 1 corresponds to the minimum and 99 to the maximum flow at which the pump can be set.
Relative flow is different for each pump. In most cases, 10 ± 5 will correspond to 1 L/min and 80 ± 5 will
correspond to 5 L/min. Also displayed on the
sampling screen
when you touch the up or down arrow
key during a run (if Screen Lock is not activated).