Reading Charge Status on Cradle LED 

The left LED on the charging cradle indicates battery status. Observe the LED steadily for > 5 seconds. 

LED Action 

Charge Status 

Red 

 
 

steady 

 
Charge in progress 

Red 

 

3 sec 

Green 

 

1 sec 

(Pattern 
repeats) 

 
Approximately 75% charged 

Green 

 
 

steady 

 
Charge completed/trickle charge 

 
Notes and Cautions 

 

Power off pump before removing battery. 

 

Use only the SKC charging cradle Cat. No. 220-800 or 220-900 for pump. 

 

Failure to follow warnings, notes, and cautions may cause injuries and voids any warranty. 

 

WARNING: Substitution of components may impair intrinsic safety. AVERTISSEMENT: La substitution de 
composants peut compromettre la Sécurité Intrinsèque. 

 

CAUTION: The battery used in this device may present a risk of fire or explosion when heated above 212 F 
(100 C) or incinerated. Replace battery with SKC Battery Pack model P75718 only. Use of another battery may 
present a risk of fire or explosion. 

 

WARNING: To prevent ignition of a hazardous atmosphere, batteries must only be changed [removed and 
replaced] in an area known to be non-hazardous. AVERTISSEMENT: Afin de p

révenir l’inflammation 

d’atmosphères dangereuses, ne changer les batteries que dans des emplacements désignés non dangereux.

 

 

Maximum charge input voltage is U

m

 = 12 V 

 

CAUTION: Risk of Fire and Burns. Do Not Disassemble, heat above 212 F (100 C), or incinerate. Keep battery 
out of reach of children and in original package until ready to use. Dispose of used batteries promptly according 
to [all state and] local recycling or waste regulations. 

 

User may replace external components such as the inlet filter, battery, protective screen cover, and/or belt clip. 
Service must be done by SKC to maintain performance and IS rating. Warranty is void if pumping compartment 
is opened by user. 
 

For more information on SKC pump lithium-ion (Li-Ion) battery packs, visit 
skcinc.com/catalog/pdf/instructions/1918.pdf 

Turning Pump Power On/Off (

Figure 1

Turn on:

 Press and hold briefly the power on/off button on the side of the pump. A 

startup screen will very briefly display all status indicators, followed by the firmware 
version (

right

) and then the set/calibrate flow screen or sampling screen showing 

elapsed time from a previous run. 

See Using the Operation Screens. 

Turn off:

 Press and hold briefly the power on/off button on the side of the pump. 

The screen will turn off.  

Note

: A non-running pump will shut off automatically after 

60 seconds of inactivity but will hold the elapsed time from the last operation until 
the timer is reset. 

Note

: The on/off button also locks and unlocks the touch pad and dims the screen during sampling (see 

Options/Modes during Sampling) 

 

 

Firmware version is displayed for 3 seconds

Summary of Contents for 220-3000

Page 1: ...th cord filter cassette holder All in One adjustable tube holder and Type A protective tube cover in a soft sided nylon carry case 100 240 V 220 3000 K3D 3 pack High Low Flow Pump Kit includes 3 pumps with Li Ion battery packs screwdriver set 3 Standard Charging Cradles power supply with cord and 3 each filter cassette holders All in One adjustable tube holders and Type A protective tube covers in...

Page 2: ... next cradle Repeat the connection to chain up to five Standard Charging Cradles Insert the connector of Multi Cradle Power Supply Cat No 220 700 into the power port on the side of the last cradle in the chain Insert the wall cube into a 100 to 240 volt wall outlet 3 Align the contacts on the bottom edge of the pump with the contacts inside the cradle and insert the pump in the cradle Repeat for e...

Page 3: ...m charge input voltage is Um 12 V CAUTION Risk of Fire and Burns Do Not Disassemble heat above 212 F 100 C or incinerate Keep battery out of reach of children and in original package until ready to use Dispose of used batteries promptly according to all state and local recycling or waste regulations User may replace external components such as the inlet filter battery protective screen cover and o...

Page 4: ...See Using the Operation Screens Using the Touch Keypad Lift the protective screen cover Use your fingertip to touch the three keys on the screen keypad Up and down arrows increase or decrease flow rate during sampling touching either arrow key briefly displays the relative flow indicator See Set Calibrate Flow Screen for more details PUMP starts and pauses stops pump touch and hold to reset elapse...

Page 5: ...pump status Green flashing Running Red flashing Flow fault Note Status LEDs will flash red green to indicate that the pump is out of flow tolerance just before entering flow fault mode and during each auto restart attempt while in flow fault mode Elapsed run time HH MM SS Battery status See Note Screen lock feature activated Flow Fault or Low Battery Fault Relative Flow A number between 1 and 99 t...

Page 6: ...he PUMP key on the screen touch keypad to return to the set calibrate flow screen 2 Prepare the calibrator See calibrator instructions 3 Set up a calibration train Figure 3 4 Using the up and down arrow keys on the pump set calibrate flow screen adjust flow until the calibrator indicates the method specified flow rate 5 Touch the PUMP key to pause stop the pump The pump is now set to the selected ...

Page 7: ...uctions use flexible tubing to connect the calibrator outlet suction port to the pump inlet 3 Set pump flow rate using the up and down arrow keys on the set calibrate flow screen to increase or decrease flow a Single tube sampling 1 5 L min b Multiple tube sampling the sum of all flows 15 Note Do not exceed 500 ml min flow rate per tube for multiple tube sampling 4 Disconnect the tubing from the p...

Page 8: ... Figure 5 Calibration Train 5 to 500 ml min for Multiple Tubes Flow adjust screw All in One Sorbent tube Suction port Pump inlet Tubing Tubing chek mate calibrator Pump chek mate calibrator Pump Dual adjustable low flow holder Sorbent tube Flow adjust screw Pump inlet CPC Tubing Tubing Suction port ...

Page 9: ...tive tube 4 Once flow is calibrated for the tube it is recommended practice to recheck the flow rate before removing the tube Any adjustment should be minimal 5 Disconnect the pump from the representative sampling medium and calibrator Proceed to Sampling Sampling Allow pump to equilibrate after moving it from one temperature extreme to another Charge pump battery completely before calibration and...

Page 10: ... that pump is turned off before removing the battery pack and that no tubing or media are attached to the pump 1 Turn the pump off by pressing the on off button 2 Remove the existing battery pack a Use a 2 5 mm hex driver Allen wrench to loosen two screws on the bottom of the battery pack housing b Pull the battery pack housing away from the pump case c If replacing the battery pack with a new Cat...

Page 11: ...pening in the belt clip and into the pump case Tighten the screw until engaged Do not tighten completely 6 Replace the screen cover a Place the screen cover and mounting block so that the two holes are aligned with the holes in the top of the belt clip Insert the two screws into the mounting block and tighten until snug b Ensure that the screen cover closes properly 7 Tighten the screw under the b...

Page 12: ...iable performance and to maintain the UL Listing for intrinsic safety Failure to do so voids any warranty Use of a repaired or rebuilt battery pack VOIDS ANY WARRANTY SKC Limited Warranty and Return Policy SKC products are subject to the SKC Limited Warranty and Return Policy which provides SKC s sole liability and the buyer s exclusive remedy To view the complete SKC Limited Warranty and Return P...

Page 13: ...l 5 of set point after calibration to desired flow Atmospheric pressure 0 3 in Hg Temperature 1 0 C Temperature ranges Operating 32 to 104 F 0 to 40 C Charging 32 to 113 F 0 to 45 C Storage 4 to 113 F 20 to 45 C Humidity ranges Operating 95 RH non condensing Storage 95 RH non condensing Altitude Corrects flow for changes in temperature 32 to 104 F 0 to 40 C and ambient pressure up to 15 000 feet 4...

Reviews: