background image

skcinc.com 

Form 37165 Rev 200217 

Page 4 

 

 
 
 

Determining Battery Charge Status 

The battery status icon at the top right of the pump display screen has four bars that decrease in number as battery 
charge is depleted. Use the table below to interpret the battery status. 

Icon Displayed 

Battery Charge Remaining 

 
Four bars 

Full battery charge, approximately 75 to 100% 

 
Three bars 

Approximately 50 to 75% 

 
Two bars 

Approximately 25 to 50% 

 
One bar 

Approximately 5 to 25% 

 
No bars 

Low battery fault is imminent. Pump will stop and power off eventually. When pump is 
powered on again, elapsed run time and fault icon will be displayed on the screen until 
pump is reset. 

See Using the Operation Screens.

 

 

Using the Touch Keypad 

Lift the protective screen cover. Use your fingertip to touch the three keys on the screen keypad:  

U

p and down arrows

 increase or decrease flow rate; during sampling, touching 

either arrow key briefly displays the relative flow indicator. 

See Set/Calibrate Flow 

Screen for more details.  

PUMP

 starts and pauses/stops pump; touch and hold to reset elapsed time. 

Using the Operation Screens: Set/Calibrate Flow and Sampling 

 
Set/Calibrate Flow Screen

  

Touch and hold the up and down arrow keys on the touch keypad to increase or decrease flow rate (

see below

) with 

pump connected to a calibrator. 

Actual flow rate is displayed on calibrator only.

  

 

 

 

 

   

 

 

 

 

 
 

 

 

 
 

 
 

 

 

Battery status

 

indicator

  

See Determining Battery Charge Status.

 

Relative flow indicator 

See Relative Flow below. 

 

Flow rate being increased 

 

Flow rate being decreased 

 

Summary of Contents for 220-3000

Page 1: ...th cord filter cassette holder All in One adjustable tube holder and Type A protective tube cover in a soft sided nylon carry case 100 240 V 220 3000 K3D 3 pack High Low Flow Pump Kit includes 3 pumps with Li Ion battery packs screwdriver set 3 Standard Charging Cradles power supply with cord and 3 each filter cassette holders All in One adjustable tube holders and Type A protective tube covers in...

Page 2: ... next cradle Repeat the connection to chain up to five Standard Charging Cradles Insert the connector of Multi Cradle Power Supply Cat No 220 700 into the power port on the side of the last cradle in the chain Insert the wall cube into a 100 to 240 volt wall outlet 3 Align the contacts on the bottom edge of the pump with the contacts inside the cradle and insert the pump in the cradle Repeat for e...

Page 3: ...m charge input voltage is Um 12 V CAUTION Risk of Fire and Burns Do Not Disassemble heat above 212 F 100 C or incinerate Keep battery out of reach of children and in original package until ready to use Dispose of used batteries promptly according to all state and local recycling or waste regulations User may replace external components such as the inlet filter battery protective screen cover and o...

Page 4: ...See Using the Operation Screens Using the Touch Keypad Lift the protective screen cover Use your fingertip to touch the three keys on the screen keypad Up and down arrows increase or decrease flow rate during sampling touching either arrow key briefly displays the relative flow indicator See Set Calibrate Flow Screen for more details PUMP starts and pauses stops pump touch and hold to reset elapse...

Page 5: ...pump status Green flashing Running Red flashing Flow fault Note Status LEDs will flash red green to indicate that the pump is out of flow tolerance just before entering flow fault mode and during each auto restart attempt while in flow fault mode Elapsed run time HH MM SS Battery status See Note Screen lock feature activated Flow Fault or Low Battery Fault Relative Flow A number between 1 and 99 t...

Page 6: ...he PUMP key on the screen touch keypad to return to the set calibrate flow screen 2 Prepare the calibrator See calibrator instructions 3 Set up a calibration train Figure 3 4 Using the up and down arrow keys on the pump set calibrate flow screen adjust flow until the calibrator indicates the method specified flow rate 5 Touch the PUMP key to pause stop the pump The pump is now set to the selected ...

Page 7: ...uctions use flexible tubing to connect the calibrator outlet suction port to the pump inlet 3 Set pump flow rate using the up and down arrow keys on the set calibrate flow screen to increase or decrease flow a Single tube sampling 1 5 L min b Multiple tube sampling the sum of all flows 15 Note Do not exceed 500 ml min flow rate per tube for multiple tube sampling 4 Disconnect the tubing from the p...

Page 8: ... Figure 5 Calibration Train 5 to 500 ml min for Multiple Tubes Flow adjust screw All in One Sorbent tube Suction port Pump inlet Tubing Tubing chek mate calibrator Pump chek mate calibrator Pump Dual adjustable low flow holder Sorbent tube Flow adjust screw Pump inlet CPC Tubing Tubing Suction port ...

Page 9: ...tive tube 4 Once flow is calibrated for the tube it is recommended practice to recheck the flow rate before removing the tube Any adjustment should be minimal 5 Disconnect the pump from the representative sampling medium and calibrator Proceed to Sampling Sampling Allow pump to equilibrate after moving it from one temperature extreme to another Charge pump battery completely before calibration and...

Page 10: ... that pump is turned off before removing the battery pack and that no tubing or media are attached to the pump 1 Turn the pump off by pressing the on off button 2 Remove the existing battery pack a Use a 2 5 mm hex driver Allen wrench to loosen two screws on the bottom of the battery pack housing b Pull the battery pack housing away from the pump case c If replacing the battery pack with a new Cat...

Page 11: ...pening in the belt clip and into the pump case Tighten the screw until engaged Do not tighten completely 6 Replace the screen cover a Place the screen cover and mounting block so that the two holes are aligned with the holes in the top of the belt clip Insert the two screws into the mounting block and tighten until snug b Ensure that the screen cover closes properly 7 Tighten the screw under the b...

Page 12: ...iable performance and to maintain the UL Listing for intrinsic safety Failure to do so voids any warranty Use of a repaired or rebuilt battery pack VOIDS ANY WARRANTY SKC Limited Warranty and Return Policy SKC products are subject to the SKC Limited Warranty and Return Policy which provides SKC s sole liability and the buyer s exclusive remedy To view the complete SKC Limited Warranty and Return P...

Page 13: ...l 5 of set point after calibration to desired flow Atmospheric pressure 0 3 in Hg Temperature 1 0 C Temperature ranges Operating 32 to 104 F 0 to 40 C Charging 32 to 113 F 0 to 45 C Storage 4 to 113 F 20 to 45 C Humidity ranges Operating 95 RH non condensing Storage 95 RH non condensing Altitude Corrects flow for changes in temperature 32 to 104 F 0 to 40 C and ambient pressure up to 15 000 feet 4...

Reviews: