2
Performance Profile
• RFI/EMI shielded
• CE marked
•
Listed for intrinsic safety (single port models only)
• Model 210-1002TX is ATEX approved
Flow Range in Constant
Flow Mode:
20 to 225 ml/min
Accuracy Variance Between
LCD Reading and Actual Flow
Rate (after calibration):
20 to 225 ml/min ± 5%
Constant Flow Compensation
Range (
inlet only
):
20 to 225 ml/min up to 20 inches water back pressure
Flow Control:
Holds constant flow to
± 5% of the set-point
Pressure Range in Constant
•
1 to 10 inches water (1.87 to 18.7 mm Hg) at
Pressure (multiple-tube
maximum flow rate of 200 ml/min
sampling) Mode (
inlet only
):
•
10 to 20 inches water (18.7 to 37.4 mm Hg) at
maximum flow rate of 100 ml/min
Constant Pressure (multiple-
tube sampling) Mode Accuracy:
Pressure reading ± 0.5 inches water (0.25 mm Hg)
Battery Charge Level Indicator:
Icon displays at full, mid, and low charge
Temperature Range:
Operating:
32 to 113 F (0 to 45 C)
Charging:
40 to 100 F (5 to 38 C)
Storage:
-4 to 113 F (-20 to 45 C)
Operating Humidity:
0 to 95% non-condensing
Protect sample pump from weather when in use outdoors.
Tubing:
Requires 1/4-in ID tubing
Run Time:
• With NiMH battery:
12 hrs at 200 ml/min up to
(dependent on sample media used)
10 inches water back pressure
12 hrs at 100 ml/min up to 20 inches water back
pressure
(Run time is 8 hrs at either setting if using a
NiCad battery.)
• Connected to charger:
Extended run time