3
Performance Profile
Timer:
1 to 9999 min display ± 1% accuracy
Flow Fault:
If the pump is unable to compensate for longer than
15 seconds due to excessive back pressure, the pump
enters FLOW FAULT mode. The pump goes into
HOLD, the fault icon appears on the display, and the
accumulated run time display is frozen and retained.
After 5 minutes in flow fault, auto-restart is attempted
every 5 minutes until flow is corrected or the user
removes the pump from flow fault mode by pressing
[
].
Battery Pack:
•
Rechargeable nickel-metal hydride (NiMH)
battery pack,
2.4 V x 1.0 Ah
•
Rechargeable nickel cadmium (NiCad) battery
pack,
2.4 V x 0.6 Ah
Charge Time:
Attached to pump:
≤ 6 hrs
Detached from pump:
16 hrs (
recommended charging
method for initial charge of a new battery pack
)
Size:
4.5 x 2.2 x 1.4 in (11.4 x 5.6 x 3.6 cm) -
see photo below
Weight:
5 oz (142 gm)
The use of a repaired or rebuilt battery pack voids any warranty and
the UL Listing for intrinsic safety.
2.2 inches
4.5 inches
1.4 inches
Newer pump models contain an NiMH
battery. An NiMH battery can be used with
older Pocket Pump models. Order NiMH
Battery Pack Cat. No. P20129-2.
Pocket Pump was designed to minimize
inherent noise. If the noise level is
distracting when worn in a shirt pocket,
clip the pump on a waist belt.
Note:
Note:
(varies with battery capacity
and level of discharge)