USER´S MANUAL – Sitmed Equipamentos Médicos Ltda
24
The basic features of the MXS 330 – ELEVEX cot, such as: Side handles, movable backrest, safety belts and casters brake are performed
according to the same instructions presented for the MRS 310 – ESSENTIAL cot, with na exception for the pantographic articulation
operation and the anti-recoil system, which accompanies the fastener system.
The MXS 330 – ELEVEX fastener system has an additional safety hook (1), called anti-recoil system. This system aims to ensure the safety
during all the operations of placing and taking the cot from the ambulance. Proceed to install the fastener system following the same
instructions described for the MRS 310 – ESSENTIAL cot. At the end of this procedure, install the safety anti-recoil hook, following as
described:
2. Place the MXS 330 – ELEVEX air-axle at the entry of the ambulance, aligned with the fastener system so that there is room for the
wheels frame to rise and fall freely. 3. Attach the safety hook (A) on the movable bar (B), taking into account the movement of the bar,
causing it to be tensioned. Mark the holes (C), remove the safety hook and the cot, drill the marked holes (be careful that the location
chosen for the hook positioning does not compromise or damage the structure of the vehicle during the drilling process) and secure the
hook with the supplied screws. 4. Note that the hook should be positioned offset from the locking turtle at approximately 10cm (D), for the
left or right.
All the movement operation and articulation of the MXS 330 – ELEVEX cot, must be performed by at least two rescuers.
1
2
3
4
Never use the stretcher without installing the additional anti-recoil safety system. Sitmed is not responsible for
accidents or damages caused from the non-installation of this component. If you have any questions during
installation, please contact our technical department immediately.