35
|
P a g e
Archimede Portable Dataloggers User Manual
rev.5 -03.12.2019-
Once a tube has been defined and the definitions of the probes you wish to use have been
configured, it is possible to perform inclinometer tests on tubes.
In case DL switches off during an inclinometer test, for example for a low battery, the readings
that have already been performed are saved. However, the inclinometer file closes, so it will be no
possible to perform other readings to complete the test. Anyway, it will be possible to delete the
partial test and repeat it in future.
Click
Read inclinometer
in the Main menu, then select the site in which the inclinometer tube is
located. The DL will show the list of the tubes defined in the site.
Select the tube you wish to use and press
ENTER
to proceed.
The DL will ask you to enter the number for the test, showing as pre-configured value the first
number available.
Enter a new number or simply press
ENTER
to proceed. You could perform up to 99 test on the
same tube.
The DL will ask for the next instrument to be used.
Select a probe and press
ENTER
.
READING INCLINOMETER TUBES