43
|
P a g e
Archimede Portable Dataloggers User Manual
rev.5 -03.12.2019-
When performing a TREX test, the screen looks as follows
:
The top blue box displays the
Current values read in mV
on both channels of the TREX probe.
When a ring is detected, both values are include in a range between 1000mV and 9000mV.
On the right the
Current Direction
(Up in our case) is shown.
The bottom blue box displays the
Current Level
of the probe inside the tube and the
Reference
value
read in the same run and at the same depth of the reference test.
At the center the current value is shown. This is displayed only when both rings have been
detected: a single value is calculated from the two channels of the T-REX. The values read on the
two channels are converted in millimeters and their difference is shown.
T-REX TEST
PROBE NAME
DATE AND TIME
MEASURE UNIT
REFERENCE VALUES
CURRENT LEVEL
TEST NAME
CURRENT GROOVE
CURRENT VALUES IN mV